Next Generation Pharmaceutical Impactor: A New Impactor for Pharmaceutical Inhaler Testing. Part III. Extension of Archival Calibration to 15 L/min
Journal of Aerosol Medicine · 2004-12 · 17 (4) · 335-343
DOI: 10.1089/jam.2004.17.335 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Aerosol Medicine, 2004-12. Verfasst von Virgil A Marple und 5 weiteren Autor:innen. Beteiligte Einrichtung: Particle Calibration Laboratory, Mechanical Engineering Department, University of Minnesota, Minneapolis, Minnesota, USA. Hauptthemen laut MEDLINE: Metered Dose Inhalers, Nebulizers and Vaporizers. Die Arbeit wurde bisher 102-mal zitiert (Crossref) und nennt 14 Literaturangaben.
Abstract (Originalsprache)
An extension of the archival calibration of the recently developed 30-100-L/min seven-stage impactor, the Next Generation Pharmaceutical Impactor (NGI), has been undertaken at 15 L/min. The NGI stage cut sizes are 0.98-14.1 microm aerodynamic diameter at this flow rate. This 15-L/min calibration was motivated by the desire to sample the entire aerosol produced by a nebulizer when tested in accordance with a new international standard developed by the Comite Européen de Normalisation (CEN), as well as the need to test various types of inhalers at flow rates lower than 30 L/min for pediatric applications. Measurements were undertaken with monodisperse oleic acid droplets in the range of 0.7-22 microm aerodynamic diameter following a procedure established in the original 30-100-L/min calibration study. The NGI was found to be effective for particle size separation at 15 L/min. Users should decide the most applicable configuration that meets their needs, based on the following recommendations: (1) the pre-separator should not normally be used, as its performance is significantly degraded by the influence of gravity, resulting in interference with stage 1; and (2) a filter should be inserted below the micro-orifice collector (MOC), as the size corresponding to 80% collection efficiency of the MOC becomes excessively large with decreasing flow rate, so that this component becomes ineffective as a means of collecting fine particles that penetrate beyond stage 7.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Particle Calibration Laboratory, Mechanical Engineering Department, University of Minnesota, Minneapolis, Minnesota, USA.
Zitieren
APA
Marple, V. A., Olson, B. A., Santhanakrishnan, K., Roberts, D. L., Mitchell, J. P., & Hudson-Curtis, B. L. (2004). Next Generation Pharmaceutical Impactor: A New Impactor for Pharmaceutical Inhaler Testing. Part III. Extension of Archival Calibration to 15 L/min. Journal of Aerosol Medicine, 17(4), 335-343. https://doi.org/10.1089/jam.2004.17.335
Vancouver
Marple VA, Olson BA, Santhanakrishnan K, Roberts DL, Mitchell JP, Hudson-Curtis BL. Next Generation Pharmaceutical Impactor: A New Impactor for Pharmaceutical Inhaler Testing. Part III. Extension of Archival Calibration to 15 L/min. Journal of Aerosol Medicine. 2004;17(4):335-343. doi:10.1089/jam.2004.17.335
BibTeX
@article{Marple2004,
author = {Marple, Virgil A and Olson, Bernard A and Santhanakrishnan, Kumaragovindhan and Roberts, Daryl L and Mitchell, Jolyon P and Hudson-Curtis, Buffy L},
title = {Next Generation Pharmaceutical Impactor: A New Impactor for Pharmaceutical Inhaler Testing. Part III. Extension of Archival Calibration to 15 L/min},
journal = {Journal of Aerosol Medicine},
year = {2004},
volume = {17},
number = {4},
pages = {335--343},
doi = {10.1089/jam.2004.17.335}
}
EUVIMED