Fachartikel in der Zeitschrift Journal of Endourology, 2023. Verfasst von Taile Jing und 15 weiteren Autor:innen. Beteiligte Einrichtung: Department of Urology, The First Affiliated Hospital, Zhejiang University School of Medicine, Hangzhou, China. Hauptthemen laut MEDLINE: Prostatectomy, Robotic Surgical Procedures. Die Arbeit wurde bisher 17-mal zitiert (Crossref) und nennt 34 Literaturangaben.
Abstract (Originalsprache)
Purpose
The aim of this study was to explore the safety and efficacy of radical prostatectomy with a novel Shurui single-port (SR-SP) robotic surgical system.
Methods
A total of 11 patients with prostate cancer were enrolled in this study. Extraperitoneal radical prostatectomy was performed using the SR-SP robotic surgical system for all patients. Clinicopathologic data, perioperative data, and short-term surgical outcomes were prospectively collected and analyzed.
Results
Of the 11 patients, the median age was 65 years (range 52-73), and the median body mass index was 22.6 kg/m2 (range 20.2-26.7). The median operative time was 229 minutes (range 194-317), and the median console time was 167 minutes (range 141-265). The median blood loss was 40 mL (range 10-120), and none of the patients required intraoperative transfusion. There was no conversion to open surgery during the operation, and no assistant ports were added. The surgeons reported a good task load rating with a National Aeronautics and Space Administration Task Load Index (NASA-TLX) score of 25.1 ± 3.3 points. The median postoperative hospital stay time was 7 days (range 4-15). There were no severe intraoperative or postoperative complications (Clavien grade ≥3). Postoperative positive surgical margin occurred in 4 (36.4%) patients. No biochemical recurrence occurred within 1 month of surgery. The continence rate was 72.7% (8/11) 1 month after surgery.
Conclusions
The new SR-SP robotic surgical system is safe, effective, flexible, and stable for application in radical prostatectomy.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Urology, The First Affiliated Hospital, Zhejiang University School of Medicine, Hangzhou, China.
- Operating Room, Department of Surgery, The First Affiliated Hospital, Zhejiang University School of Medicine, Hangzhou, China.
Zitieren
APA
Jing, T., Peng, D., Yao, X., Ye, S., Zhao, Q., Ni, M., Shao, J., Xu, X., He, A., Liang, Z., Lai, C., Qi, H., Meng, H., Wang, P., Wang, S., & Xia, D. (2023). Single-Port Robot–Assisted Radical Prostatectomy with the Novel Shurui Single-Port Robotic Surgical System. Journal of Endourology. https://doi.org/10.1089/end.2023.0185
Vancouver
Jing T, Peng D, Yao X, Ye S, Zhao Q, Ni M, et al. Single-Port Robot–Assisted Radical Prostatectomy with the Novel Shurui Single-Port Robotic Surgical System. Journal of Endourology. 2023. doi:10.1089/end.2023.0185
BibTeX
@article{Jing2023,
author = {Jing, Taile and Peng, Ding and Yao, Xiaolin and Ye, Sunyi and Zhao, Qianqian and Ni, Meng and Shao, Jing and Xu, Xin and He, Anbang and Liang, Zhen and Lai, Chong and Qi, Honggang and Meng, Hongzhou and Wang, Ping and Wang, Shuo and Xia, Dan},
title = {Single-Port Robot–Assisted Radical Prostatectomy with the Novel Shurui Single-Port Robotic Surgical System},
journal = {Journal of Endourology},
year = {2023},
doi = {10.1089/end.2023.0185}
}
EUVIMED