Accuracy of Continuous Glucose Monitoring in Adults with Type 1 Diabetes Admitted to Hospital: A Real-World Multicenter Observational Study

Ray Wang iD, Mervyn Kyi, Brintha Krishnamoorthi, Jason Tjahyadi, Ailie Connell, Cherie Chiang, Debra Renouf, Rahul Barmanray iD, Spiros Fourlanos

Diabetes Technology & Therapeutics · 2025

DOI: 10.1089/dia.2024.0604 ↗

Kurz gefasst

Beobachtungsstudie in der Zeitschrift Diabetes Technology & Therapeutics, 2025. Verfasst von Ray Wang und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Diabetes & Endocrinology, The Royal Melbourne Hospital, Parkville, Australia. Hauptthemen laut MEDLINE: Diabetes Mellitus, Type 1, Blood Glucose, Blood Glucose Self-Monitoring. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 42 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Continuous glucose monitoring (CGM) use in people with type 1 diabetes (T1D) is revolutionizing management. Use of CGM in hospital is poised to transform care, however routine use is not currently recommended due to lack of accuracy validation in acute care, including in people with T1D. We aimed to determine real-world CGM accuracy in hospitalized adults with T1D. Materials and Methods: In this multicenter retrospective observational study, we compared CGM interstitial fluid glucose with reference blood glucose (capillary/whole-blood point-of-care [POC], blood gas [GAS]) in adults with T1D requiring multiday admissions during 2020–2023 across three health services in Australia. Patients requiring dialysis or admitted under pediatric/obstetric/palliative care/psychiatry units were excluded. CGM accuracy was assessed by comparison with time-matched (±5 min) reference glucose measures, utilizing median absolute relative difference (ARD), mean ARD (MARD), and consensus error grid (CEG) analysis.

Results

In total, 2,199 CGM-reference glucose pairs from 214 admissions (146 patients) were assessed. Overall, mean (SD) ARD was 12.8% (13.1) and median (IQR) ARD was 9.4% (3.7–17.7). MARD for CGM-POC pairs was 12.3%; MARD for CGM-GAS pairs was 14.3%. In CEG analysis, 99.3% of glucose pairs were within zones A/B. Accuracy was lower in critical care compared with noncritical care wards (MARD 16.1% vs. 12.0%, P < 0.001).

Conclusions

In this real-world multicenter study, CGM glucose agreed well with reference blood glucose, suggesting modern CGM devices could be safely and effectively used in hospitalized adults with T1D. Further prospective studies of CGM accuracy with newer generation devices across different scenarios will further elucidate inpatient CGM accuracy and safety.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Diabetes Mellitus, Type 1Blood GlucoseBlood Glucose Self-MonitoringHumansRetrospective StudiesFemaleAdultMaleMiddle AgedHospitalizationAustraliaYoung AdultReproducibility of ResultsExtracellular FluidContinuous Glucose Monitoring

Stichwörter der Autor:innen

CGMCGMSaccuracycontinuous glucose monitoringhospitalizationinpatient

Substanzen

Blood Glucose

Autor:innen und Institutionen

  • Department of Diabetes & Endocrinology, The Royal Melbourne Hospital, Parkville, Australia.
  • Department of Medicine (Royal Melbourne Hospital), The University of Melbourne, Parkville, Australia.
  • Australian Centre for Accelerating Diabetes Innovations (ACADI), The University of Melbourne, Parkville, Australia.
  • Department of Endocrinology & Diabetes, Eastern Health, Box Hill, Australia.
  • Department of Diabetes & Endocrinology, Frankston, Australia.
  • Department of Chemical Pathology, Eastern Health, Box Hill, Australia.

ORCID-Profile: Ray Wang, Rahul Barmanray

Zitieren

APA

Wang, R., Kyi, M., Krishnamoorthi, B., Tjahyadi, J., Connell, A., Chiang, C., Renouf, D., Barmanray, R., & Fourlanos, S. (2025). Accuracy of Continuous Glucose Monitoring in Adults with Type 1 Diabetes Admitted to Hospital: A Real-World Multicenter Observational Study. Diabetes Technology & Therapeutics. https://doi.org/10.1089/dia.2024.0604

Vancouver

Wang R, Kyi M, Krishnamoorthi B, Tjahyadi J, Connell A, Chiang C, et al. Accuracy of Continuous Glucose Monitoring in Adults with Type 1 Diabetes Admitted to Hospital: A Real-World Multicenter Observational Study. Diabetes Technology & Therapeutics. 2025. doi:10.1089/dia.2024.0604

BibTeX
@article{Wang2025,
  author = {Wang, Ray and Kyi, Mervyn and Krishnamoorthi, Brintha and Tjahyadi, Jason and Connell, Ailie and Chiang, Cherie and Renouf, Debra and Barmanray, Rahul and Fourlanos, Spiros},
  title = {Accuracy of Continuous Glucose Monitoring in Adults with Type 1 Diabetes Admitted to Hospital: A Real-World Multicenter Observational Study},
  journal = {Diabetes Technology & Therapeutics},
  year = {2025},
  doi = {10.1089/dia.2024.0604}
}
Kostenlos starten →