Accuracy of a Continuous Glucose Monitor in the Intensive Care Unit: A Proposed Accuracy Standard and Calibration Protocol for Inpatient Use
Diabetes Technology & Therapeutics · 2024
DOI: 10.1089/dia.2024.0074 ↗Kurz gefasst
Studienprotokoll in der Zeitschrift Diabetes Technology & Therapeutics, 2024. Verfasst von Sewon A Bann und 6 weiteren Autor:innen. Beteiligte Einrichtung: Division of Endocrinology, University of British Columbia, Vancouver, Canada. Hauptthemen laut MEDLINE: Intensive Care Units, Blood Glucose, Inpatients. Die Arbeit wurde bisher 19-mal zitiert (Crossref) und nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Themen: Intensivpflege
Abstract (Originalsprache)
Background and Aims: Guidelines now recommend inpatient continuous glucose monitor (CGM) use with confirmatory blood glucose measurements. However, the Food and Drug Administration has not yet officially approved CGM for inpatient use in large part because its accuracy has not been established in this setting. We tested the accuracy of the Dexcom G6 (G6) in 28 adults on an insulin infusion in a medical–surgical intensive care unit with 1064 matched CGM and arterial point-of-care pairs. Methods: The participants were on average 57.29 (SD 2.39) years, of whom 13 had a prior diagnosis of diabetes and 14 were admitted for a surgical diagnosis. The first 19 participants received the G6 without calibration and had a mean absolute relative difference (MARD) of 13.19% (IQR 5.11, 19.03) across 659 matched pairs, which just meets the critical care expert recommendation of MARD
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Division of Endocrinology, University of British Columbia, Vancouver, Canada.
- Department of Biological Sciences, Simon Fraser University, Burnaby, Canada.
- Deanery of Molecular, Genetic and Population Health Sciences, University of Edinburgh, Edinburgh, UK.
- Division of Critical Care, University of British Columbia, Vancouver, Canada.
ORCID-Profile: Sewon A Bann
Zitieren
APA
Bann, S. A., Hercus, J. C., Atkins, P., Alkhairy, A., Loyal, J. P., Sekhon, M., & Thompson, D. J. (2024). Accuracy of a Continuous Glucose Monitor in the Intensive Care Unit: A Proposed Accuracy Standard and Calibration Protocol for Inpatient Use. Diabetes Technology & Therapeutics. https://doi.org/10.1089/dia.2024.0074
Vancouver
Bann SA, Hercus JC, Atkins P, Alkhairy A, Loyal JP, Sekhon M, et al. Accuracy of a Continuous Glucose Monitor in the Intensive Care Unit: A Proposed Accuracy Standard and Calibration Protocol for Inpatient Use. Diabetes Technology & Therapeutics. 2024. doi:10.1089/dia.2024.0074
BibTeX
@article{Bann2024,
author = {Bann, Sewon A and Hercus, Jess C and Atkins, Paul and Alkhairy, Areej and Loyal, Jackson P and Sekhon, Mypinder and Thompson, David J},
title = {Accuracy of a Continuous Glucose Monitor in the Intensive Care Unit: A Proposed Accuracy Standard and Calibration Protocol for Inpatient Use},
journal = {Diabetes Technology & Therapeutics},
year = {2024},
doi = {10.1089/dia.2024.0074}
}
EUVIMED