First Use of Open-Source Automated Insulin Delivery AndroidAPS in Full Closed-Loop Scenario: Pancreas4ALL Randomized Pilot Study
Diabetes Technology & Therapeutics · 2023
DOI: 10.1089/dia.2022.0562 ↗Kurz gefasst
Randomisierte kontrollierte Studie in der Zeitschrift Diabetes Technology & Therapeutics, 2023. Verfasst von Lenka Petruzelkova und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Pediatrics, Motol University Hospital, 2nd Faculty of Medicine, Charles University, Prague, Czech Republic. Hauptthemen laut MEDLINE: Insulin, Diabetes Mellitus, Type 1. Die Arbeit wurde bisher 39-mal zitiert (Crossref) und nennt 25 Literaturangaben.
Abstract (Originalsprache)
Objective
We evaluated the safety and feasibility of open-source automated insulin delivery AndroidAPS in adolescents and young adults with type 1 diabetes (T1D) and compared its efficacy in three different scenarios: hybrid closed loop (HCL) with meal boluses, meal announcement only (MA), and full closed loop (FCL). Research Design and Methods: In an open-label, prospective, randomized crossover trial ( clinicaltrials.gov NCT04835350), 16 adolescents with T1D (10 females) with mean age of 17 years (range 15–20), glycated hemoglobin 56 mmol/mol (range 43–75), and mean duration of diabetes 5.9 years (9–15) underwent three distinct 3-day periods of camp living, comparing the above-mentioned scenarios of AndroidAPS. We used modified and locked version of AndroidAPS 3.1.03, which was called Pancreas4ALL for study purposes. The order of MA and FCL periods was assigned randomly. The primary endpoints were feasibility and safety of the system represented by percentage of time of glucose control by the system and time in hypoglycemia below 3 mmol/L.
Results
The glycemia was controlled by the system 95% time of the study and the proportion of time below 3 mmol/L did not exceed 1% over the whole study period (0.72%). The HCL scenario reached significantly higher percentage of time below 3 mmol/L (HCL 1.05% vs. MA 0.0% vs. FCL 0.0%; P = 0.05) compared to other scenarios. No difference was observed among the scenarios in the percentage of time between 3.9 and 10 mmol/L (HCL 83.3% vs. MA 79.85% vs. FCL 81.03%, P = 0.58) corresponding to mean glycemia (HCL 6.65 mmol/L vs. MA 7.34 mmol/L vs. FCL 7.05 mmol/L, P = 0.28). No difference was observed in the mean daily dose of insulin or in the daily carbohydrate intake. No serious adverse event occurred during the study period.
Conclusions
Our pilot study showed that FCL might be a realistic mode of treatment for people with T1D.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Pediatrics, Motol University Hospital, 2nd Faculty of Medicine, Charles University, Prague, Czech Republic.
- IT Department, CLOSED LOOP Systems and Sysop, Prague, Czech Republic.
ORCID-Profile: Lenka Petruzelkova
Zitieren
APA
Petruzelkova, L., Neuman, V., Plachy, L., Kozak, M., Obermannova, B., Kolouskova, S., Pruhova, S., & Sumnik, Z. (2023). First Use of Open-Source Automated Insulin Delivery AndroidAPS in Full Closed-Loop Scenario: Pancreas4ALL Randomized Pilot Study. Diabetes Technology & Therapeutics. https://doi.org/10.1089/dia.2022.0562
Vancouver
Petruzelkova L, Neuman V, Plachy L, Kozak M, Obermannova B, Kolouskova S, et al. First Use of Open-Source Automated Insulin Delivery AndroidAPS in Full Closed-Loop Scenario: Pancreas4ALL Randomized Pilot Study. Diabetes Technology & Therapeutics. 2023. doi:10.1089/dia.2022.0562
BibTeX
@article{Petruzelkova2023,
author = {Petruzelkova, Lenka and Neuman, Vit and Plachy, Lukas and Kozak, Milos and Obermannova, Barbora and Kolouskova, Stanislava and Pruhova, Stepanka and Sumnik, Zdenek},
title = {First Use of Open-Source Automated Insulin Delivery AndroidAPS in Full Closed-Loop Scenario: Pancreas4ALL Randomized Pilot Study},
journal = {Diabetes Technology & Therapeutics},
year = {2023},
doi = {10.1089/dia.2022.0562}
}
EUVIMED