Electromagnetic simulation of graphene-based microtransistor arrays in MRI using an accelerated Huygens’ box sub-gridding framework
Medical Sensors & Imaging · 2026
DOI: 10.1088/2977-8425/aea620 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Sensors & Imaging, 2026. Verfasst von Suchit Kumar und 8 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Engineering and Physical Sciences Research Council.
Abstract (Originalsprache)
Implementing electrophysiological recordings within an MRI environment is challenging due to complex interactions between recording probes and MRI-generated fields, which can affect both safety and data quality. This study aims to develop and evaluate a hybrid electromagnetic (EM) simulation framework for efficient and accurate assessment of such interactions.
Methods
A hybrid EM strategy integrating the Huygens'Box (HB) method with sub-gridding was implemented in an FDTD solver (Sim4Life). RF coil models for mouse and rat head were simulated with and without intracortical (IC) and epicortical (EC) graphene-based micro-transistor arrays. Three-dimensional multi-layered probe models were reconstructed from two-dimensional layouts, and transmit field ($B_{1}^{+}$), electric field ($E$), and specific absorption rate (SAR) distributions were evaluated. Performance was benchmarked against conventional full-wave multi-port (MP) simulations using Bland-Altman analysis and voxel-wise percentage differences.
Results
HB simulations reduced computational time by approximately 70-80%, while preserving spatial patterns of $|B_{1}^{+}|$, $|E|$, and SAR, including transmit-field symmetry and localized high-field regions. Deviations from MP were minimal for $|B_{1}^{+}|$ (median $\Delta$% 0.02-0.07% in mice, -3.7% to -1.7% in rats) and modest for $|E|$ and SAR, with absolute SAR values remaining well below human safety limits. Graphene-based arrays produced negligible effects on RF transmission and SAR deposition.
Conclusion
The HB approach enables computationally efficient, high-resolution evaluation of EM interactions involving microscopic probes in MRI environments, supporting simulations that are otherwise impractical with full-wave MP modeling.
Förderung
- Engineering and Physical Sciences Research Council
Zitieren
APA
Kumar, S., Ramírez, A. L., Flaherty, S., Guimera, A., Eulate, N. A. D., Kostarelos, K., Dickie, B., Wykes, R. C., & Lemieux, L. (2026). Electromagnetic simulation of graphene-based microtransistor arrays in MRI using an accelerated Huygens’ box sub-gridding framework. Medical Sensors & Imaging. https://doi.org/10.1088/2977-8425/aea620
Vancouver
Kumar S, Ramírez AL, Flaherty S, Guimera A, Eulate NAD, Kostarelos K, et al. Electromagnetic simulation of graphene-based microtransistor arrays in MRI using an accelerated Huygens’ box sub-gridding framework. Medical Sensors & Imaging. 2026. doi:10.1088/2977-8425/aea620
BibTeX
@article{Kumar2026,
author = {Kumar, Suchit and Ramírez, Alejandro Labastida and Flaherty, Samuel and Guimera, Anton and Eulate, Nerea Alvarez de and Kostarelos, Kostas and Dickie, Ben and Wykes, Rob C and Lemieux, Louis},
title = {Electromagnetic simulation of graphene-based microtransistor arrays in MRI using an accelerated Huygens’ box sub-gridding framework},
journal = {Medical Sensors & Imaging},
year = {2026},
doi = {10.1088/2977-8425/aea620}
}
EUVIMED