Comparisons of seismocardiographic sensor positions for VO 2 max estimations

Sofie Braun Svensen iD, Julie Kathrine Dittmann Andreasen iD, Joan Hentze, Emil Korsgaard iD, Rasmus Gundorff Sæderup iD, Peter Søgaard, Ahmad Agam, Samuel Emil Schmidt iD

Biomedical Physics & Engineering Express · 18.11.2026

DOI: 10.1088/2057-1976/aec3ab ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biomedical Physics & Engineering Express, online erschienen am 18.11.2026. Verfasst von Sofie Braun Svensen und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Health Science and Technology, Aalborg University, AAU, Aalborg, Denmark. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

VO 2 max is a key health indicator of cardiorespiratory fitness and overall physical performance. VO 2 max can be estimated using seismocardiography (SCG). SCG can be measured at multiple anatomical locations. The xiphoid process (XP) has primarily been used for VO 2 max estimation, and the comparative performance of alternative sensor locations has not been evaluated. The aim of this study was to determine whether SCG sensor locations other than the XP produce systematically different VO 2 max estimates when the same estimation algorithm is applied.

The study included 13 healthy male participants. SCG signals were sampled from a 4 × 4 grid of accelerometers, resulting in 16 different sensor locations across the chest, including the XP. VO 2 max was estimated using a previously validated non-exercise SCG algorithm based on fiducial SCG features and demographic variables. Differences in estimated VO 2 max between each sensor location and the XP reference were analysed using a linear mixed-effects model, and Pearson's correlation coefficients were calculated to assess the relationship between sample locations.
The analysis showed that seven sensor locations differed significantly from the XP reference, with mean deviations ranging from 1.108 to 2.626 ml·kg⁻¹·min⁻¹. All correlations were greater than 0.849, indicating high consistency between estimates. Bland–Altman analyses demonstrated small mean biases ranging from −2.47 to 2.03 ml·kg⁻¹·min⁻¹ and no evidence of proportional bias for any sensor location, although the limits of agreement indicated some variability at the individual level. VO 2 max estimates showed the strongest agreement with the XP reference at the right thorax, left sternum, and lower left thorax overlying the cardiac silhouette compared with the remaining measurement locations. 

These findings suggest that several alternative SCG sensor locations provide VO 2 max estimates comparable to the standard XP location, whereas a limited number of locations produce systematic differences and may require location-specific correction before clinical implementation.

Schlagwörter

Stichwörter der Autor:innen

Cardiorespiratory fitnessSeismoFitSeismocardiographySensor positions for VO2max estimationsVO2maxVO2max placementsensor placement

Autor:innen und Institutionen

  • Department of Health Science and Technology, Aalborg University, AAU, Aalborg, Denmark.
  • Department of Cardiology, Aalborg University Hospital, Aalborg, Denmark.
  • Department of Clinical Medicine, Aalborg University, AAU, Aalborg, Denmark.

ORCID-Profile: Sofie Braun Svensen, Julie Kathrine Dittmann Andreasen, Emil Korsgaard, Rasmus Gundorff Sæderup, Samuel Emil Schmidt

Zitieren

APA

Svensen, S. B., Andreasen, J. K. D., Hentze, J., Korsgaard, E., Sæderup, R. G., Søgaard, P., Agam, A., & Schmidt, S. E. (2026). Comparisons of seismocardiographic sensor positions for VO 2 max estimations. Biomedical Physics & Engineering Express. https://doi.org/10.1088/2057-1976/aec3ab

Vancouver

Svensen SB, Andreasen JKD, Hentze J, Korsgaard E, Sæderup RG, Søgaard P, et al. Comparisons of seismocardiographic sensor positions for VO 2 max estimations. Biomedical Physics & Engineering Express. 2026. doi:10.1088/2057-1976/aec3ab

BibTeX
@article{Svensen2026,
  author = {Svensen, Sofie Braun and Andreasen, Julie Kathrine Dittmann and Hentze, Joan and Korsgaard, Emil and Sæderup, Rasmus Gundorff and Søgaard, Peter and Agam, Ahmad and Schmidt, Samuel Emil},
  title = {Comparisons of seismocardiographic sensor positions for VO 2 max estimations},
  journal = {Biomedical Physics & Engineering Express},
  year = {2026},
  doi = {10.1088/2057-1976/aec3ab}
}
Kostenlos starten →