Modeling of novel converging beam radiotherapy system in Geant4 Monte Carlo simulation
Biomedical Physics & Engineering Express · 17.11.2026
DOI: 10.1088/2057-1976/aec2f8 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biomedical Physics & Engineering Express, online erschienen am 17.11.2026. Verfasst von Dongyoen Lee und 8 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Convergent Radiotherapy and Radiosurgery Ltd..
Abstract (Originalsprache)
Purpose
A Geant4-based model was created to simulate radiation transport within a novel radiotherapy system developed by Convergent Radiotherapy and Radiosurgery (Convergent RnR; Haifa, Israel). This treatment device incorporates a Bragg-diffraction X-ray lens to focus X-rays from a conventional X-ray tube towards a small but finite focal spot with quasi-monoenergetic photons.
Methods
The simulation model included an X-ray tube and a CRnR focusing lens consisting of single-crystal aluminum tiles. To validate the Geant4 model of this device, transverse and longitudinal in-air and in-water dose distributions from the converging beam field were calculated and compared to radiochromic film measurements.
Results
The implemented Monte Carlo (MC) simulation model predicted a mean photon energy of 53.96 keV and showed agreement within 0.5% when compared with the SpekPy calculation. Both the simulation and experiment produced an attenuation characteristics of 3.2 mm Al. In-water tissue maximum ratios showed percentage differences ≤3.03% at depths of 5, 10, 30, 40, 50, 60, and 70 mm. Exceptions were observed at 20, 80, and 85 mm, likely due to film positioning uncertainty. A gamma analysis of in-water scanning dose distributions demonstrated a >96% pass rate for the 2%/2 mm criterion and >99% for the 3%/3 mm criterion across all focal depths.
Conclusions
In this study, CRnR’s converging beam radiotherapy system was implemented in Geant4 to demonstrate the device operation and characterize its physical and dosimetric outputs. The MC simulation implemented in this work successfully demonstrated converging dose distributions with Bragg-diffraction physics, which were in good agreement with radiochromic film measurements. This simulation model may be used to further refine the converging beam system design for clinical implementation, including validation of the treatment planning system.
Förderung
- Convergent Radiotherapy and Radiosurgery Ltd.
Zitieren
APA
Lee, D., Reiazi, R., Bartkoski, D. A., Schori, A., Insley, B. A., Fru, L. C., Tailor, R., Kleckner, M., & Salehpour, M. R. (2026). Modeling of novel converging beam radiotherapy system in Geant4 Monte Carlo simulation. Biomedical Physics & Engineering Express. https://doi.org/10.1088/2057-1976/aec2f8
Vancouver
Lee D, Reiazi R, Bartkoski DA, Schori A, Insley BA, Fru LC, et al. Modeling of novel converging beam radiotherapy system in Geant4 Monte Carlo simulation. Biomedical Physics & Engineering Express. 2026. doi:10.1088/2057-1976/aec2f8
BibTeX
@article{Lee2026,
author = {Lee, Dongyoen and Reiazi, Reza and Bartkoski, Dirk Alan and Schori, Aviad and Insley, Benjamin Abraham and Fru, Leonard Che and Tailor, Ramesh and Kleckner, Michael and Salehpour, Mohammad R},
title = {Modeling of novel converging beam radiotherapy system in Geant4 Monte Carlo simulation},
journal = {Biomedical Physics & Engineering Express},
year = {2026},
doi = {10.1088/2057-1976/aec2f8}
}
EUVIMED