Development of a cost-effective bladder-prostate phantom model to simulate the effects of minimally invasive surgical treatments (MIST) for benign prostatic hyperplasia (BPH) and its validation by pressure-flow study
Biomedical Physics & Engineering Express · 10.11.2026
DOI: 10.1088/2057-1976/aec01e ↗Kurz gefasst
Fachartikel in der Zeitschrift Biomedical Physics & Engineering Express, online erschienen am 10.11.2026. Verfasst von Joy Narayan Chakraborty und 2 weiteren Autor:innen. Beteiligte Einrichtung: Consultant urologist, Hayat Hospital, Guwahati, India. Hauptthemen laut MEDLINE: Minimally Invasive Surgical Procedures, Prostatic Hyperplasia, Urinary Bladder, Prostate. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Development of a cost-effective Bladder-Prostate Phantom model to simulate the effects of minimally invasive surgical treatments (MIST) for benign prostatic hyperplasia (BPH) and its validation by pressure-flow studyAbstractWe developed and bench-tested an economical bladder-prostate training construct to simulate bladder outlet obstruction and its relief during non-thermal minimally invasive interventions. The aim was to create a simple, locally fabricable platform to practise urethral implantation and quantify its urodynamic effects. The system combined a paediatric silicone AmbuBag (≈500 mL, wall thickness 3.6 mm, Shore 78 Sh0) as a compressible reservoir, a funnel-shaped neck, a collapsible silicone urethral tube (internal diameter 10 mm, wall thickness 0.95 mm, Shore 65 Sh0), and polyurethane foam balls (Shore 25 Sh00) to represent prostatic lobes compressing a 5-cm urethral segment. Using a multichannel urodynamic device, we recorded pressure-flow parameters in three configurations: obstruction alone, obstruction with an intraurethral coil-spring implant, and obstruction with a traction implant. Under obstruction, an infusion rate of 75 mL/min yielded a voided volume of 351 mL, an end-fill pressure of 8 cm H₂O, an opening detrusor pressure of 35 cm H₂O, a peak flow (Qmax) of 7.6 mL/s, a PdetQmax (detrusor pressure at peak flow) of 102 cm H₂O, and a bladder outlet obstruction index (BOOI) of 87.0, consistent with severe outflow resistance. With the coil-spring implant, peak flow increased to 31.8 mL/s, with PdetQmax reduced to 11 cm H₂O and the maximum detrusor pressure to 78 cm H₂O. With the traction implant, peak flow increased again to 36.7 mL/s, with an opening pressure of 84 cm H₂O, a PdetQmax of 14 cm H₂O, and a maximum detrusor pressure of 109 cm H₂O. These changes demonstrate improved pressure-flow performance, supporting this construct as a practical platform for device testing and skills training.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Consultant urologist, Hayat Hospital, Guwahati, India.
- Department of Mechanical Engineering, IITG, Guwahati, India.
ORCID-Profile: Joy Narayan Chakraborty
Zitieren
APA
Chakraborty, J. N., Kanagaraj, S., & Muthu, N. (2026). Development of a cost-effective bladder-prostate phantom model to simulate the effects of minimally invasive surgical treatments (MIST) for benign prostatic hyperplasia (BPH) and its validation by pressure-flow study. Biomedical Physics & Engineering Express. https://doi.org/10.1088/2057-1976/aec01e
Vancouver
Chakraborty JN, Kanagaraj S, Muthu N. Development of a cost-effective bladder-prostate phantom model to simulate the effects of minimally invasive surgical treatments (MIST) for benign prostatic hyperplasia (BPH) and its validation by pressure-flow study. Biomedical Physics & Engineering Express. 2026. doi:10.1088/2057-1976/aec01e
BibTeX
@article{Chakraborty2026,
author = {Chakraborty, Joy Narayan and Kanagaraj, Subramani and Muthu, Nelson},
title = {Development of a cost-effective bladder-prostate phantom model to simulate the effects of minimally invasive surgical treatments (MIST) for benign prostatic hyperplasia (BPH) and its validation by pressure-flow study},
journal = {Biomedical Physics & Engineering Express},
year = {2026},
doi = {10.1088/2057-1976/aec01e}
}
EUVIMED