EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Feasibility of a depth camera-based center of mass feedback system for closed-loop functional electrical stimulation therapy and visual feedback balance training

Derrick Lim iD, William Pei iD, Jae W Lee iD, Kristin E Musselman iD, Kei Masani iD

Biomedical Physics & Engineering Express · 01.08.2026 · 12 (4) · 045038

DOI: 10.1088/2057-1976/ae89bc ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biomedical Physics & Engineering Express, online erschienen am 23.07.2026. Verfasst von Derrick Lim und 4 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Biomedical Engineering, University of Toronto, Toronto, Canada. Hauptthemen laut MEDLINE: Postural Balance, Electric Stimulation Therapy, Feedback, Sensory, Spinal Cord Injuries. Die Arbeit nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Natural Sciences and Engineering Research Council of Canada, CIHR/NSERC/SSHRC Collaborative Health Research Project.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Individuals with incomplete spinal cord injury are at an increased risk of falls due to impaired balance control. Our group previously developed a closed-loop visual feedback balance training (VFBT) system combined with functional electrical stimulation (FES) therapy to provide standing balance therapy (FES + VFBT system). However, its clinical use has been limited by the reliance on force plates, which are costly and not widely available. The purpose of this study was to develop and evaluate a clinically accessible FES + VFBT system that uses a depth camera in place of a force plate to provide center of mass (COM) feedback. Ten neurologically intact participants (6 males, 4 female) completed three sets of four balance exercises. The COM estimates from the depth camera-based system showed very good correlation with the force plate-based system ( r AP > 0.85; r ML > 0.7) and low root mean squared error, RMSE, (RMSE AP < 0.009 m; RMSE ML < 0.017 m). The stimulation profiles derived from the COM estimates from the depth camera-based system also closely matched those from the force plate system (RMSE < 1.8 mA). These results demonstrate that the proposed depth camera-based FES + VFBT system performs comparably to the original system, offering a more accessible solution for balance therapy in clinical settings.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Postural BalanceElectric Stimulation TherapyFeedback, SensorySpinal Cord InjuriesHumansFemaleMaleAdultFeasibility StudiesAccidental Falls

Stichwörter der Autor:innen

depth camerafunctional electrical stimulationstanding balancevisual feedback balance training

Autor:innen und Institutionen

  • Institute of Biomedical Engineering, University of Toronto, Toronto, Canada.
  • KITE-Toronto Rehabilitation Institute, University Health Network, Toronto, Canada.
  • Department of Neurology, Oregon Health & Science University, Portland, USA.
  • Rehabilitation Sciences Institute, University of Toronto, Toronto, Canada.
  • Department of Physical Therapy, University of Toronto, Toronto, Canada.

ORCID-Profile: Derrick Lim, William Pei, Jae W Lee, Kristin E Musselman, Kei Masani

Förderung

  • Natural Sciences and Engineering Research Council of Canada
  • CIHR/NSERC/SSHRC Collaborative Health Research Project

Zitieren

APA

Lim, D., Pei, W., Lee, J. W., Musselman, K. E., & Masani, K. (2026). Feasibility of a depth camera-based center of mass feedback system for closed-loop functional electrical stimulation therapy and visual feedback balance training. Biomedical Physics & Engineering Express, 12(4), 045038. https://doi.org/10.1088/2057-1976/ae89bc

Vancouver

Lim D, Pei W, Lee JW, Musselman KE, Masani K. Feasibility of a depth camera-based center of mass feedback system for closed-loop functional electrical stimulation therapy and visual feedback balance training. Biomedical Physics & Engineering Express. 2026;12(4):045038. doi:10.1088/2057-1976/ae89bc

BibTeX
@article{Lim2026,
  author = {Lim, Derrick and Pei, William and Lee, Jae W and Musselman, Kristin E and Masani, Kei},
  title = {Feasibility of a depth camera-based center of mass feedback system for closed-loop functional electrical stimulation therapy and visual feedback balance training},
  journal = {Biomedical Physics & Engineering Express},
  year = {2026},
  volume = {12},
  number = {4},
  pages = {045038},
  doi = {10.1088/2057-1976/ae89bc}
}
Kostenlos starten →