Validation of 12-lead Smartphone-based spandan ultra 12L electrocardiogram durations and amplitudes analysis as compared to the standard 12-lead electrocardiogram

Chandra Mohan, Kunal Gururani, Anurag Rawat, Yogendra Singh, Nitin Chandola iD, Deeksha Agarwal, Sengar Yashwardhan Pratap Singh iD

Biomedical Physics & Engineering Express · 01.04.2026 · 12 (2) · 025062

DOI: 10.1088/2057-1976/ae596f ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Biomedical Physics & Engineering Express, online erschienen am 09.04.2026. Verfasst von Chandra Mohan und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Himalayan Institute of Medical Sciences, Swami Rama Himalayan University, Dehradun, Uttarakhand, India. Hauptthemen laut MEDLINE: Smartphone, Electrocardiography. Die Arbeit nennt 23 Literaturangaben. Gefördert von: Sunfox Technologies Pvt. Ltd.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Electrocardiography (ECG) is a valuable tool for diagnosing cardiac disorders, with a 12-lead ECG being the gold standard. However, the traditional 12-lead ECG machine is hospital-based, which restricts access. The Smartphone-based 12-Lead ECG (Spandan Ultra 12L ECG) is a portable, simultaneous 12-lead ECG device designed for real-time cardiac monitoring. This study compares its accuracy with a Standard 12-lead ECG system (Philips PageWriter ECG).

Objective

To evaluate the accuracy and reliability of the Smartphone-based 12-Lead ECG device in measuring ECG interval, amplitude, and duration measurements as compared to the Standard 12-lead ECG.

Methods

A prospective, single-blinded, cross-sectional study was conducted in a hospital ECG room in Dehradun, India. 405 participants aged ≥ 20 years in ECG recordings using the both the ECG machine. After applying the exclusion criteria, 75 participants were excluded, and 330 participants’ ECG recordings were analyzed. Statistical analyses included paired t-tests to compare significant differences, Pearson correlation coefficients (PCC) to assess linear association, and Bland-Altman (BA) analysis to evaluate the agreement between the two devices.

Results

Twelve-lead Standard (n = 330) and Smartphone-based 12-Lead (n = 330) ECG waveforms from 330 participants (mean age 52.14 years); 169 (51.21%) males were analyzed. No statistically significant differences were found between both the ECGs for all measured parameters, with p-values ranging from 0.090 for P-wave duration to 0.965 for Q-wave amplitude. PCC showed high linear correlation between the both the ECGs across ECG intervals (0.818–0.938), amplitudes (>0.92), and durations (0.700–0.944). BA analysis confirmed minimal bias, with limits of agreement (LOA) within clinically acceptable ranges.

Conclusion

The Smartphone-based 12-Lead ECG device showed high accuracy and reliability in the evaluation of ECG parameters, as compared to the Standard 12-lead ECG system. Its portability, smartphone compatibility, and ease of use make it a promising diagnostic tool for regular clinical applications and timely ECG assessment.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

SmartphoneElectrocardiographyHumansMaleReproducibility of ResultsProspective StudiesCross-Sectional StudiesFemaleMiddle AgedAdultSingle-Blind MethodAgedSignal Processing, Computer-Assisted

Stichwörter der Autor:innen

10 s ECG12-lead ECGclinical validationmedical deviceportable ECGprehospital emergency careremote patient monitoring

Autor:innen und Institutionen

  • Department of Cardiology, Himalayan Institute of Medical Sciences, Swami Rama Himalayan University, Dehradun, Uttarakhand, India.
  • Department of Cardiology, Max Super Speciality Hospitals, Dehradun, Uttarakhand, India.
  • Research and development, Department of Clinical Research, Sunfox Technologies, Dehradun, Uttarakhand, India.

ORCID-Profile: Nitin Chandola, Sengar Yashwardhan Pratap Singh

Förderung

  • Sunfox Technologies Pvt. Ltd

Zitieren

APA

Mohan, C., Gururani, K., Rawat, A., Singh, Y., Chandola, N., Agarwal, D., & Pratap Singh, S. Y. (2026). Validation of 12-lead Smartphone-based spandan ultra 12L electrocardiogram durations and amplitudes analysis as compared to the standard 12-lead electrocardiogram. Biomedical Physics & Engineering Express, 12(2), 025062. https://doi.org/10.1088/2057-1976/ae596f

Vancouver

Mohan C, Gururani K, Rawat A, Singh Y, Chandola N, Agarwal D, et al. Validation of 12-lead Smartphone-based spandan ultra 12L electrocardiogram durations and amplitudes analysis as compared to the standard 12-lead electrocardiogram. Biomedical Physics & Engineering Express. 2026;12(2):025062. doi:10.1088/2057-1976/ae596f

BibTeX
@article{Mohan2026,
  author = {Mohan, Chandra and Gururani, Kunal and Rawat, Anurag and Singh, Yogendra and Chandola, Nitin and Agarwal, Deeksha and Pratap Singh, Sengar Yashwardhan},
  title = {Validation of 12-lead Smartphone-based spandan ultra 12L electrocardiogram durations and amplitudes analysis as compared to the standard 12-lead electrocardiogram},
  journal = {Biomedical Physics & Engineering Express},
  year = {2026},
  volume = {12},
  number = {2},
  pages = {025062},
  doi = {10.1088/2057-1976/ae596f}
}
Kostenlos starten →