Simulation of lung volume and SPECT count errors due to mismatch between SPECT and CT during free-breathing in lung perfusion scintigraphy

Simulation von Lungenvolumen- und SPECT-Count-Fehlern durch Fehlanpassung zwischen SPECT und CT bei freier Atmung in der Lungenszintigraphie

Yuta Nojima iD, Yoshihiro Yamazaki iD

Biomedical Physics & Engineering Express · 01.02.2026 · 12 (1) · 017001

DOI: 10.1088/2057-1976/ae2ebb ↗

Auf einen Blick

Die Simulation verglich freiatmende und CT-Atemanhalt-Aufnahmen mit synchronisierter Akquisition. Die Fehlanpassung zwischen SPECT und CT zeigte keinen statistisch signifikanten Einfluss auf die quantitative Genauigkeit.

Fragestellung

Wie wirken sich Fehlanpassungen zwischen SPECT- und CT-Atemphase bei freier Atmung auf Lungenvolumen und SPECT-Counts in der Lungenszintigraphie aus?

Methode

Simulation mit 4D-CT-Bildern aus 10 Atemphasen; digitale Phantomkonstruktion und SPECT-Rekonstruktion. Vergleich von synchronisierten, freiatmenden und CT-Atemanhalt-Szenarien.

Ergebnisse

Bei Inspiration waren die Lungenvolumina geringer, bei Expiration höher als in synchronisierten Bildern. Gegenüber respiratorisch gegatelter SPECT ergaben sich bei freier Atmung Fehler von etwa 2,5 % beim Lungenvolumen und 1,2 % bei den SPECT-Counts.

Schlussfolgerung der Autor:innen

Die Autor:innen schlussfolgern, dass die Phasenfehlanpassung zwischen SPECT und CT die quantitative Genauigkeit nicht statistisch signifikant beeinflusste und die freie Atmung die diagnostische Fähigkeit nur minimal beeinträchtigte.

Kernaussagen

  • Freie Atmung: etwa 2,5 % Fehler beim Lungenvolumen
  • Freie Atmung: etwa 1,2 % Fehler bei SPECT-Counts
  • Keine statistisch signifikante Verbesserung durch die μMAP-Phase

Deutsche Zusammenfassung: KI-gestützt aus dem Original-Abstract erstellt; alle Zahlen wurden automatisch mit dem Original abgeglichen. Maßgeblich ist der Original-Abstract.

Kurz gefasst

Fachartikel in der Zeitschrift Biomedical Physics & Engineering Express, online erschienen am 05.01.2026. Verfasst von Yuta Nojima und 1 weiteren Autor:innen. Beteiligte Einrichtung: Graduate School of Health Sciences, Niigata University, 2-746 Asahimachi-dori, Chuo-ku, Niigata 951-8518, Japan. Hauptthemen laut MEDLINE: Lung, Tomography, Emission-Computed, Single-Photon, Perfusion Imaging, Respiration. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Japan Society for the Promotion of Science.

Auf einen BlickAbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Respiratory phase mismatch between single-photon emission computed tomography (SPECT) and computed tomography (CT) acquisition phases presents a challenge in lung perfusion scintigraphy using SPECT/CT. This study simulated lung volume and SPECT counts changes under free-breathing and breath-hold CT conditions compared to respiratory-synchronized acquisition. Chest 4D-CT images, divided into 10 respiratory phases, were used to generate lung, soft tissue, liver, and bone regions for each phase. A digital phantom was constructed via image processing using ImageJ. SPECT images were generated from these phantoms by employing the Prominence Processor to simulate projection data and reconstruct images. Simulations included a ‘synchronized image,’ where both SPECT and μMAP for attenuation correction were created in the same phase; a ‘free-breathing image,’ combining a free-breathing SPECT and μMAP; and a ‘CT breath-hold image,’ using phase-specific μMAPs with the free-breathing SPECT image for attenuation correction. Lung volumes and SPECT counts in the free-breathing and CT breath-hold images were compared with those in the synchronized image. By analyzing the relative errors caused by differences in the μMAPs, the study evaluated the impact of mismatch between SPECT and CT phases. Results indicated that lung volumes appeared reduced during inspiration and increased during expiration compared with synchronized images. No significant difference in the relative error was observed between the free-breathing and CT breath-hold images. Our findings revealed that in the quantitative evaluation of lung perfusion SPECT, varying the μ-map phase during free-breathing acquisition did not result in a significant improvement, suggesting that the mismatch between SPECT and CT had no statistically significant effect on quantitative accuracy. Compared with respiratory-gated SPECT, free-breathing acquisitions introduced potential errors of approximately 2.5% in lung volume measurement and 1.2% in SPECT counts. However, these errors were within acceptable tolerance limits for clinical diagnosis, indicating that free-breathing acquisition had minimal effects on diagnostic capability.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

LungTomography, Emission-Computed, Single-PhotonPerfusion ImagingRespirationTomography, X-Ray ComputedSingle Photon Emission Computed Tomography Computed TomographyHumansPhantoms, ImagingImage Processing, Computer-AssistedComputer Simulation

Stichwörter der Autor:innen

SPECT/CTattenuation correctionlung volumerespiratory phase mismatchsimulation study

Autor:innen und Institutionen

  • Graduate School of Health Sciences, Niigata University, 2-746 Asahimachi-dori, Chuo-ku, Niigata 951-8518, Japan.

ORCID-Profile: Yuta Nojima, Yoshihiro Yamazaki

Förderung

  • Japan Society for the Promotion of Science

Zitieren

APA

Nojima, Y., & Yamazaki, Y. (2026). Simulation of lung volume and SPECT count errors due to mismatch between SPECT and CT during free-breathing in lung perfusion scintigraphy. Biomedical Physics & Engineering Express, 12(1), 017001. https://doi.org/10.1088/2057-1976/ae2ebb

Vancouver

Nojima Y, Yamazaki Y. Simulation of lung volume and SPECT count errors due to mismatch between SPECT and CT during free-breathing in lung perfusion scintigraphy. Biomedical Physics & Engineering Express. 2026;12(1):017001. doi:10.1088/2057-1976/ae2ebb

BibTeX
@article{Nojima2026,
  author = {Nojima, Yuta and Yamazaki, Yoshihiro},
  title = {Simulation of lung volume and SPECT count errors due to mismatch between SPECT and CT during free-breathing in lung perfusion scintigraphy},
  journal = {Biomedical Physics & Engineering Express},
  year = {2026},
  volume = {12},
  number = {1},
  pages = {017001},
  doi = {10.1088/2057-1976/ae2ebb}
}
Kostenlos starten →