Design and implementation of a pulsed LED system as low-intensity light source for measuring SiPMs parameters
Journal of Instrumentation · 01.07.2026 · 21 (07) · C07013
DOI: 10.1088/1748-0221/21/07/c07013 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Instrumentation, online erschienen am 13.07.2026. Verfasst von C.C. Radulescu und 2 weiteren Autor:innen. Die Arbeit nennt 11 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Low-intensity light sources, down to single-photon emissions, are indispensable for characterizing the photon detection devices coupled to the readout electronics. Current work present a circuit for driving light emitting diodes in pulsed mode to emit a small number of photons needed to measure specific parameters for silicon photomultipliers or equivalent photodetectors. Low cost off-the-shelf components have been used through the implementation based on a microcontroller operating two delay lines that drive a combinational logic circuit to obtain a short width pulse triggering a ultra-fast gate driver, further switching a gallium nitride transistor with the LED in high-side configuration. The LTSpice simulations were consistent with timing performance of the practical system allowing to increment the pulse width in steps of 0.25 ns from 4 ns to 60 ns with a dynamic repetition rate starting from 1 Hz up to 200 Hz. An experimental cross-check was made between the parameters of a single-channel SiPM determined with a benchmark laser source versus the values obtained using the pulsed LED system, and they were in good agreement. Either in controlled laboratory conditions or at irradiation facilities, this system met the requirements to extract the SiPMs key parameters starting with breakdown voltage, gain and photodetection efficiency at different wavelengths, including aging tests.
Zitieren
APA
Radulescu, C., Cojocariu, L., & Lupusoru, V. (2026). Design and implementation of a pulsed LED system as low-intensity light source for measuring SiPMs parameters. Journal of Instrumentation, 21(07), C07013. https://doi.org/10.1088/1748-0221/21/07/c07013
Vancouver
Radulescu C, Cojocariu L, Lupusoru V. Design and implementation of a pulsed LED system as low-intensity light source for measuring SiPMs parameters. Journal of Instrumentation. 2026;21(07):C07013. doi:10.1088/1748-0221/21/07/c07013
BibTeX
@article{Radulescu2026,
author = {Radulescu, C.C. and Cojocariu, L.N. and Lupusoru, V.},
title = {Design and implementation of a pulsed LED system as low-intensity light source for measuring SiPMs parameters},
journal = {Journal of Instrumentation},
year = {2026},
volume = {21},
number = {07},
pages = {C07013},
doi = {10.1088/1748-0221/21/07/c07013}
}
EUVIMED