EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

NeuroTest: a benchtop testbed for evaluating sensing-capable electrophysiology and neurostimulation systems

Morrow Jk iD, H Cho iD, M Benjaber iD, Denison Tj iD, Widge As iD, Herron Ja iD

Journal of Neural Engineering · 01.12.2026 · 23 (6) · 066003

DOI: 10.1088/1741-2552/aeb82d ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Neural Engineering, online erschienen am 04.11.2026. Verfasst von Morrow Jk und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Psychiatry and Behavioral Sciences, University of Minnesota, Minneapolis, MN, Unites States of America. Hauptthemen laut MEDLINE: Implantable Neurostimulators, Electrodes, Implanted. Die Arbeit nennt 136 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIMH NIH HHS, NINDS NIH HHS, National Institute of Neurological Disorders and Stroke.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective . The development of increasingly complex implantable neuromodulation devices requires extensive benchtop validation and testing prior to in vivo implementation. Our goal is to enable this in vitro testing by creating a benchtop system that is accessible, easy-to-use, and functions reliably across academic and industry environments, reducing discordant results that arise from discordant approaches. Furthermore, such a system paves the way for standardizing benchtop characterization of future implantable systems by establishing a repeatable and comparable methodology for validating these devices and their capabilities. Approach. We describe the NeuroTest Board (NTB), a low-cost, open-source device that combines a low-noise current source waveform generator with a high-precision data acquisition system. We provide information on its architecture, operation, and performance as an initial benchmark of the potential capabilities of the system. Main results. We demonstrate the use of the NTB to assess electrophysiology systems in two separate applications. In one case, we assess a commonly used laboratory system (the RHD2132 digitizing headstage and Open Ephys data acquisition system), while the second case characterizes an investigational neuromodulation device in development for human use (the CorTec Brain InterChange). Following this initial sensing characterization, we explore stimulation-related artifacts that would hamper in vivo data collection. With the NTB, we develop and evaluate artifact mitigation strategies for each system, thus demonstrating an example application of this benchtop testbed. Significance. The ability to test neuromodulation devices against a common benchmark will enable faster development of novel therapeutics by increasing inter-institutional reliability and decreasing troubleshooting time.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Implantable NeurostimulatorsElectrodes, ImplantedEquipment DesignAnimalsHumansReproducibility of Results

Stichwörter der Autor:innen

deep brain stimulationelectrophysiologyneuromodulation

Autor:innen und Institutionen

  • Department of Psychiatry and Behavioral Sciences, University of Minnesota, Minneapolis, MN, Unites States of America.
  • Department of Neurological Surgery, University of Washington, Seattle, WA, Unites States of America.
  • Institute of Biomedical Engineering, University of Oxford, Oxford, United Kingdom.

ORCID-Profile: Morrow Jk, H Cho, M Benjaber, Denison Tj, Widge As, Herron Ja

Förderung

  • NIMH NIH HHS
  • NINDS NIH HHS
  • National Institute of Neurological Disorders and Stroke
  • NIHR Oxford Biomedical Research Centre
  • National Institue of Mental Health
  • Medical Research Council

Zitieren

APA

Jk, M., Cho, H., Benjaber, M., Tj, D., As, W., & Ja, H. (2026). NeuroTest: a benchtop testbed for evaluating sensing-capable electrophysiology and neurostimulation systems. Journal of Neural Engineering, 23(6), 066003. https://doi.org/10.1088/1741-2552/aeb82d

Vancouver

Jk M, Cho H, Benjaber M, Tj D, As W, Ja H. NeuroTest: a benchtop testbed for evaluating sensing-capable electrophysiology and neurostimulation systems. Journal of Neural Engineering. 2026;23(6):066003. doi:10.1088/1741-2552/aeb82d

BibTeX
@article{Jk2026,
  author = {Jk, Morrow and Cho, H and Benjaber, M and Tj, Denison and As, Widge and Ja, Herron},
  title = {NeuroTest: a benchtop testbed for evaluating sensing-capable electrophysiology and neurostimulation systems},
  journal = {Journal of Neural Engineering},
  year = {2026},
  volume = {23},
  number = {6},
  pages = {066003},
  doi = {10.1088/1741-2552/aeb82d}
}
Kostenlos starten →