171 Yb + optical clock with 2.2 × 10 − 18 systematic uncertainty and absolute frequency measurements
Metrologia · 01.08.2024 · 61 (4) · 045001
DOI: 10.1088/1681-7575/ad53cd ↗Kurz gefasst
Fachartikel in der Zeitschrift Metrologia, online erschienen am 17.06.2024. Verfasst von A Tofful und 11 weiteren Autor:innen. Die Arbeit wurde bisher 43-mal zitiert (Crossref) und nennt 47 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Engineering and Physical Sciences Research Council, Science and Technology Facilities Council, European Metrology Programme for Innovation and Research.
Abstract (Originalsprache)
A full evaluation of the uncertainty budget for the ytterbium ion optical clock at the National Physical Laboratory (NPL) was performed on the electric octupole (E3) 2 S 1 / 2 → 2 F 7 / 2 transition. The total systematic frequency shift was measured with a fractional standard systematic uncertainty of 2.2 × 10 − 18 . Furthermore, the absolute frequency of the E3 transition of the 171 Yb + ion was measured between 2019 and 2023 via a link to International Atomic Time (TAI) and against the local caesium fountain NPL-CsF2. The absolute frequencies were measured with fractional standard uncertainties between 3.7 × 10 − 16 and 1.1 × 10 − 15 , and all were in agreement with the 2021 BIPM recommended frequency.
Förderung
- Engineering and Physical Sciences Research Council
- Science and Technology Facilities Council
- European Metrology Programme for Innovation and Research
- Department for Business, Energy and Industrial Strategy, UK Government
Zitieren
APA
Tofful, A., Baynham, C. F. A., Curtis, E. A., Parsons, A. O., Robertson, B. I., Schioppo, M., Tunesi, J., Margolis, H. S., Hendricks, R. J., Whale, J., Thompson, R. C., & Godun, R. M. (2024). 171 Yb + optical clock with 2.2 × 10 − 18 systematic uncertainty and absolute frequency measurements. Metrologia, 61(4), 045001. https://doi.org/10.1088/1681-7575/ad53cd
Vancouver
Tofful A, Baynham CFA, Curtis EA, Parsons AO, Robertson BI, Schioppo M, et al. 171 Yb + optical clock with 2.2 × 10 − 18 systematic uncertainty and absolute frequency measurements. Metrologia. 2024;61(4):045001. doi:10.1088/1681-7575/ad53cd
BibTeX
@article{Tofful2024,
author = {Tofful, A and Baynham, C F A and Curtis, E A and Parsons, A O and Robertson, B I and Schioppo, M and Tunesi, J and Margolis, H S and Hendricks, R J and Whale, J and Thompson, R C and Godun, R M},
title = {171 Yb + optical clock with 2.2 × 10 − 18 systematic uncertainty and absolute frequency measurements},
journal = {Metrologia},
year = {2024},
volume = {61},
number = {4},
pages = {045001},
doi = {10.1088/1681-7575/ad53cd}
}
EUVIMED