Ventilation heterogeneity under mechanical ventilation via mask during anesthetic induction: a randomized trial comparing two ventilation strategies

Li Yang iD, Lingling Gao, Congxia Pan, Lili Pan, Zhanqi Zhao iD, Jun Zhang iD

Physiological Measurement · 31.08.2026 · 47 (8) · 085021

DOI: 10.1088/1361-6579/ae8d7c ↗

Kurz gefasst

Randomisierte kontrollierte Studie in der Zeitschrift Physiological Measurement, online erschienen am 18.08.2026. Verfasst von Li Yang und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anesthesiology, Fudan University Shanghai Cancer Center; Department of Oncology, Shanghai Medical College, Fudan University, Shanghai 2…. Hauptthemen laut MEDLINE: Respiration, Artificial, Masks, Anesthesia. Die Arbeit nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Shanghai Municipal Science and Technology Committee.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective. Low tidal volume, limited driving pressure and positive end expiratory pressure (PEEP) are well-accepted lung-protected ventilation strategies. The study aimed to compare driving pressure targeted strategy (PTS) and low tidal volume + PEEP strategy (VPS) in term of ventilation heterogeneity under mechanical ventilation via mask during anesthetic induction. Approach . Patients undergoing anesthesia induction were randomly assign to either PTS or VPS groups. An inspiration pressure of 15 cm H 2 O and zero PEEP was used in PTS group. A tidal volume of 8 ml kg −1 predicted body weight and a PEEP of 5 cm H 2 O was used in VPS group. The regional lung ventilation was monitored by electrical impedance tomography (EIT). Expiratory tidal volume ( TV e), driving pressure, hemodynamics and PaO 2 /FiO 2 , and EIT-derived ventilation homogeneity (global inhomogeneity index, GI ), the ratio of left/right lung ventilation distribution were calculated. Main results. Finally, 30 patients with PTS and 25 with VPS were analyzed. Compared with spontaneous ventilation, mechanical ventilation via mask caused significant changes in the GI (0.40 ± 0.04 vs 0.48 ± 0.08, P < 0.05). When compared with PTS, VPS was associated with improved GI (0.52 ± 0.09 vs 0.42 ± 0.03, PTS vs VPS, P < 0.001, power value 0.955), lower TV e (625.3 ± 192.7 vs 429.1 ± 48.8 ml, P < 0.001), lower driving pressure (15 ± 0 vs 10.5 ± 1.5 cm H 2 O, P < 0.001), and slightly higher but not statistically different PaO 2 /FiO 2 (370.7 ± 90.1 vs 391.1 ± 48.9 mm Hg), P = 0.380). Significance . Mask mechanical ventilation with VPS may improve the lung ventilation inhomogeneity during anesthetic induction when compared with PTS.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Respiration, ArtificialMasksAnesthesiaHumansFemaleMaleTidal VolumePositive-Pressure RespirationMiddle AgedElectric ImpedanceAdult

Stichwörter der Autor:innen

anesthetic inductiondriving pressure targetedelectrical impedance tomographyglobal inhomogeneity indexpositive end-expiratory pressuretidal volume

Autor:innen und Institutionen

  • Department of Anesthesiology, Fudan University Shanghai Cancer Center; Department of Oncology, Shanghai Medical College, Fudan University, Shanghai 200032, People's Republic of China.
  • Department of Critical Care Medicine, The First Affiliated Hospital of Guangzhou Medical University, Guangzhou, People's Republic of China.
  • School of Biomedical Engineering, Guangzhou Medical University, Guangzhou, People's Republic of China.
  • Department of Critical Care Medicine, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences, Beijing, People's Republic of China.
  • Institute of Technical Medicine, Furtwangen University, Villingen-Schwenningen, Germany.

ORCID-Profile: Li Yang, Zhanqi Zhao, Jun Zhang

Förderung

  • Shanghai Municipal Science and Technology Committee

Zitieren

APA

Yang, L., Gao, L., Pan, C., Pan, L., Zhao, Z., & Zhang, J. (2026). Ventilation heterogeneity under mechanical ventilation via mask during anesthetic induction: a randomized trial comparing two ventilation strategies. Physiological Measurement, 47(8), 085021. https://doi.org/10.1088/1361-6579/ae8d7c

Vancouver

Yang L, Gao L, Pan C, Pan L, Zhao Z, Zhang J. Ventilation heterogeneity under mechanical ventilation via mask during anesthetic induction: a randomized trial comparing two ventilation strategies. Physiological Measurement. 2026;47(8):085021. doi:10.1088/1361-6579/ae8d7c

BibTeX
@article{Yang2026,
  author = {Yang, Li and Gao, Lingling and Pan, Congxia and Pan, Lili and Zhao, Zhanqi and Zhang, Jun},
  title = {Ventilation heterogeneity under mechanical ventilation via mask during anesthetic induction: a randomized trial comparing two ventilation strategies},
  journal = {Physiological Measurement},
  year = {2026},
  volume = {47},
  number = {8},
  pages = {085021},
  doi = {10.1088/1361-6579/ae8d7c}
}
Kostenlos starten →