Simulation study on the dosimetric characteristics of focused electron beams with different energies for FLASH radiotherapy
Physics in Medicine & Biology · 28.06.2026 · 71 (12) · 125025
DOI: 10.1088/1361-6560/ae752f ↗Kurz gefasst
Fachartikel in der Zeitschrift Physics in Medicine & Biology, online erschienen am 24.06.2026. Verfasst von Xiong Tan und 9 weiteren Autor:innen. Beteiligte Einrichtung: The Institute for Advanced Studies, Wuhan University, Wuhan 430072, People's Republic of China. Hauptthemen laut MEDLINE: Electrons, Monte Carlo Method. Die Arbeit nennt 39 Literaturangaben. Gefördert von: Translational Medicine and Interdisciplinary Research Joint Fund of Zhongnan Hospital of Wuhan University, Major Science and Technology Project of Hubei Province.
Abstract (Originalsprache)
Objective. Focused high-energy electron beams are a promising option for FLASH radiotherapy of deep-seated tumors because beam focusing may improve dose concentration at depth while reducing unnecessary irradiation of surrounding normal tissues. This study aimed to investigate the dosimetric characteristics of focused electron beams for FLASH radiotherapy and to provide insight into electron beam energy selection. Approach. An electron beam focusing system was developed, with the beam transport line optimized using the transport code. The dosimetric performance of 1 GeV, 500 MeV, and 200 MeV electron beams passing through this system was systematically simulated in a water phantom using the Monte-Carlo code FLUKA. The dose distributions of beams with different energies were quantitatively compared under the same optical configuration. Main results. Among the investigated beam energies, the 1 GeV electron beam exhibited the strongest focusing capability and achieved the minimum full width at half maximum (FWHM) at the focal point, offering unique advantages for spatially fractionated radiotherapy. The 500 MeV beam showed only a limited increase in focal size and FWHM compared with the 1 GeV beam, whereas the 200 MeV beam exhibited relatively poor focusing performance in deep regions. Consequently, the 500 MeV electron beam may represent a viable substitute for the 1 GeV beam in specific clinical or engineering environments, as it provides a favorable balance between dosimetric performance and implementation cost. Significance. The systematic analysis of the dosimetric characteristics of electron beams at different energies provided in this study can serve as a valuable reference for related research and for the future clinical implementation of FLASH radiotherapy based on high energy electron beams.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- The Institute for Advanced Studies, Wuhan University, Wuhan 430072, People's Republic of China.
- Advanced Light Source Research Center, Wuhan University, Wuhan 430072, People's Republic of China.
- Radiotherapy Center Zhongnan Hospital of Wuhan University, Wuhan 430071, People's Republic of China.
ORCID-Profile: Xiong Tan, Chaofan An, Yuan Chen, Jike Wang, Yuancun Nie
Förderung
- Translational Medicine and Interdisciplinary Research Joint Fund of Zhongnan Hospital of Wuhan University
- Major Science and Technology Project of Hubei Province
Zitieren
APA
Tan, X., An, C., Qin, R., Dai, Z., Li, J., Chen, Y., Wang, J., Jiang, D., Liu, H., & Nie, Y. (2026). Simulation study on the dosimetric characteristics of focused electron beams with different energies for FLASH radiotherapy. Physics in Medicine & Biology, 71(12), 125025. https://doi.org/10.1088/1361-6560/ae752f
Vancouver
Tan X, An C, Qin R, Dai Z, Li J, Chen Y, et al. Simulation study on the dosimetric characteristics of focused electron beams with different energies for FLASH radiotherapy. Physics in Medicine & Biology. 2026;71(12):125025. doi:10.1088/1361-6560/ae752f
BibTeX
@article{Tan2026,
author = {Tan, Xiong and An, Chaofan and Qin, Ruoyu and Dai, Zeyi and Li, Jia and Chen, Yuan and Wang, Jike and Jiang, Dazhen and Liu, Hui and Nie, Yuancun},
title = {Simulation study on the dosimetric characteristics of focused electron beams with different energies for FLASH radiotherapy},
journal = {Physics in Medicine & Biology},
year = {2026},
volume = {71},
number = {12},
pages = {125025},
doi = {10.1088/1361-6560/ae752f}
}
EUVIMED