Comparing energy-integrating detector and photon-counting detector-based breast cone beam CTs for microcalcification detection via Monte Carlo simulation
Physics in Medicine & Biology · 28.03.2026 · 71 (6) · 065006
DOI: 10.1088/1361-6560/ae50ca ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Physics in Medicine & Biology, online erschienen am 23.03.2026. Verfasst von Ahad Ollah Ezzati und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University, Baltimore, MD, United States of America. Hauptthemen laut MEDLINE: Monte Carlo Method, Photons, Cone-Beam Computed Tomography, Calcinosis. Die Arbeit nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Institute of Biomedical Imaging and Bioengineering, National Cancer Institute.
Abstract (Originalsprache)
Objective. Microcalcification ( µ Calc) detection plays an important role in breast cancer screening. Electronic noise in energy-integrating detectors (EIDs) is the major challenge for this task in current breast cone-beam CT (bCBCT) due to the tight dose constraint for breast imaging. bCBCT with a photon counting detector (PCD) can potentially offer a higher spatial resolution and lower noise. This study performed a direct comparison of bCBCTs with the two detector types via GPU-based Monte Carlo (MC) simulation. Approach. We employed Virtual Clinical Trial for Regulatory Evaluation toolkit to generate a realistic breast phantom with a 0.25 mm 3 voxel size, 80% fat fraction and 14 cm diameter. We considered a bCBCT system with a 60 kV x-ray source filtered with 0.3 mm Cu and detector response functions for PCD and EID. A total of 360 projections were simulated with a total number of 3.15 × 10 12 photons, corresponding to ∼4 mGy mean glandular dose, comparable to a two-view mammography. We modified our GPU-based MC simulation code to incorporate analytical descriptions of µ Calcs of spherical shapes with diameters ranging from 0.1 to 0.4 mm, in 0.1 mm increments, into the voxelized phantom. A nichrome wire with 0.07 mm diameter was simulated to calculate the modulation transfer functions (MTFs). bCBCT images were reconstructed with the Feldkamp–Davis–Kress algorithm, and image quality and µ Calc detection performance were evaluated. Main results. EID-bCBCT had more profound image noise due to electronic noise. The image intensity standard deviations estimated within a region of interest were 0.055 cm −1 for EID-bCBCT and 0.038 cm −1 for PCD-bCBCT, respectively. µ Calcs and breast anatomy such as ligaments were more visible in the PCD-bCBCT images. The 10% MTF cutoffs were 5.5 and 9.5 lp mm −1 for EID-bCBCT and PCD-bCBCT, respectively. Contrast-to-noise ratio ranged in 1.20–9.13 for EID-bCBCT and 3.07–14.74 for PCD-bCBCT, depending on µ Calc sizes. Significance. We compared EID- and PCD-based bCBCT for µ Calc detection using GPU-based MC simulations in a clinically realistic setting. Our results demonstrate a potential advantage of PCD-bCBCT for this detection task.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University, Baltimore, MD, United States of America.
- Division of Diagnostic Imaging Physics, Department of Radiology, Massachusetts General Hospital, 55 Fruit Street, Boston, MA, United States of America.
ORCID-Profile: Xiaoyu Hu, Miao Qi, Youfang Lai, Kai Yang, Xun Jia
Förderung
- National Institute of Biomedical Imaging and Bioengineering
- National Cancer Institute
Zitieren
APA
Ollah Ezzati, A., Hu, X., Qi, M., Lai, Y., Zhong, Y., Yang, K., & Jia, X. (2026). Comparing energy-integrating detector and photon-counting detector-based breast cone beam CTs for microcalcification detection via Monte Carlo simulation. Physics in Medicine & Biology, 71(6), 065006. https://doi.org/10.1088/1361-6560/ae50ca
Vancouver
Ollah Ezzati A, Hu X, Qi M, Lai Y, Zhong Y, Yang K, et al. Comparing energy-integrating detector and photon-counting detector-based breast cone beam CTs for microcalcification detection via Monte Carlo simulation. Physics in Medicine & Biology. 2026;71(6):065006. doi:10.1088/1361-6560/ae50ca
BibTeX
@article{OllahEzzati2026,
author = {Ollah Ezzati, Ahad and Hu, Xiaoyu and Qi, Miao and Lai, Youfang and Zhong, Yuncheng and Yang, Kai and Jia, Xun},
title = {Comparing energy-integrating detector and photon-counting detector-based breast cone beam CTs for microcalcification detection via Monte Carlo simulation},
journal = {Physics in Medicine & Biology},
year = {2026},
volume = {71},
number = {6},
pages = {065006},
doi = {10.1088/1361-6560/ae50ca}
}
EUVIMED