Comparing energy-integrating detector and photon-counting detector-based breast cone beam CTs for microcalcification detection via Monte Carlo simulation

Ahad Ollah Ezzati, Xiaoyu Hu iD, Miao Qi iD, Youfang Lai iD, Yuncheng Zhong, Kai Yang iD, Xun Jia iD

Physics in Medicine & Biology · 28.03.2026 · 71 (6) · 065006

DOI: 10.1088/1361-6560/ae50ca ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Physics in Medicine & Biology, online erschienen am 23.03.2026. Verfasst von Ahad Ollah Ezzati und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University, Baltimore, MD, United States of America. Hauptthemen laut MEDLINE: Monte Carlo Method, Photons, Cone-Beam Computed Tomography, Calcinosis. Die Arbeit nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Institute of Biomedical Imaging and Bioengineering, National Cancer Institute.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective. Microcalcification ( µ Calc) detection plays an important role in breast cancer screening. Electronic noise in energy-integrating detectors (EIDs) is the major challenge for this task in current breast cone-beam CT (bCBCT) due to the tight dose constraint for breast imaging. bCBCT with a photon counting detector (PCD) can potentially offer a higher spatial resolution and lower noise. This study performed a direct comparison of bCBCTs with the two detector types via GPU-based Monte Carlo (MC) simulation. Approach. We employed Virtual Clinical Trial for Regulatory Evaluation toolkit to generate a realistic breast phantom with a 0.25 mm 3 voxel size, 80% fat fraction and 14 cm diameter. We considered a bCBCT system with a 60 kV x-ray source filtered with 0.3 mm Cu and detector response functions for PCD and EID. A total of 360 projections were simulated with a total number of 3.15 × 10 12 photons, corresponding to ∼4 mGy mean glandular dose, comparable to a two-view mammography. We modified our GPU-based MC simulation code to incorporate analytical descriptions of µ Calcs of spherical shapes with diameters ranging from 0.1 to 0.4 mm, in 0.1 mm increments, into the voxelized phantom. A nichrome wire with 0.07 mm diameter was simulated to calculate the modulation transfer functions (MTFs). bCBCT images were reconstructed with the Feldkamp–Davis–Kress algorithm, and image quality and µ Calc detection performance were evaluated. Main results. EID-bCBCT had more profound image noise due to electronic noise. The image intensity standard deviations estimated within a region of interest were 0.055 cm −1 for EID-bCBCT and 0.038 cm −1 for PCD-bCBCT, respectively. µ Calcs and breast anatomy such as ligaments were more visible in the PCD-bCBCT images. The 10% MTF cutoffs were 5.5 and 9.5 lp mm −1 for EID-bCBCT and PCD-bCBCT, respectively. Contrast-to-noise ratio ranged in 1.20–9.13 for EID-bCBCT and 3.07–14.74 for PCD-bCBCT, depending on µ Calc sizes. Significance. We compared EID- and PCD-based bCBCT for µ Calc detection using GPU-based MC simulations in a clinically realistic setting. Our results demonstrate a potential advantage of PCD-bCBCT for this detection task.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Monte Carlo MethodPhotonsCone-Beam Computed TomographyCalcinosisBreastHumansPhantoms, ImagingFemale

Stichwörter der Autor:innen

Monte Carlobreast CTcone beam CTphoton counting

Autor:innen und Institutionen

  • Department of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University, Baltimore, MD, United States of America.
  • Division of Diagnostic Imaging Physics, Department of Radiology, Massachusetts General Hospital, 55 Fruit Street, Boston, MA, United States of America.

ORCID-Profile: Xiaoyu Hu, Miao Qi, Youfang Lai, Kai Yang, Xun Jia

Förderung

  • National Institute of Biomedical Imaging and Bioengineering
  • National Cancer Institute

Zitieren

APA

Ollah Ezzati, A., Hu, X., Qi, M., Lai, Y., Zhong, Y., Yang, K., & Jia, X. (2026). Comparing energy-integrating detector and photon-counting detector-based breast cone beam CTs for microcalcification detection via Monte Carlo simulation. Physics in Medicine & Biology, 71(6), 065006. https://doi.org/10.1088/1361-6560/ae50ca

Vancouver

Ollah Ezzati A, Hu X, Qi M, Lai Y, Zhong Y, Yang K, et al. Comparing energy-integrating detector and photon-counting detector-based breast cone beam CTs for microcalcification detection via Monte Carlo simulation. Physics in Medicine & Biology. 2026;71(6):065006. doi:10.1088/1361-6560/ae50ca

BibTeX
@article{OllahEzzati2026,
  author = {Ollah Ezzati, Ahad and Hu, Xiaoyu and Qi, Miao and Lai, Youfang and Zhong, Yuncheng and Yang, Kai and Jia, Xun},
  title = {Comparing energy-integrating detector and photon-counting detector-based breast cone beam CTs for microcalcification detection via Monte Carlo simulation},
  journal = {Physics in Medicine & Biology},
  year = {2026},
  volume = {71},
  number = {6},
  pages = {065006},
  doi = {10.1088/1361-6560/ae50ca}
}
Kostenlos starten →