Evaluation of proton range differences in photon-counting and dual-energy computed tomography across imaging doses and anthropomorphic phantom sizes

Didier Lustermans iD, Gabriel Paiva Fonseca iD, Gloria Vilches-Freixas iD, Ilaria Rinaldi, Eva Hoeijmakers iD, Cécile Rlpn Jeukens iD, Katia Parodi iD, Guillaume Landry iD, Vicki Trier Taasti iD, Frank Verhaegen iD

Physics in Medicine & Biology · 28.03.2026 · 71 (6) · 065008

DOI: 10.1088/1361-6560/ae50c8 ↗

Kurz gefasst

Evaluationsstudie in der Zeitschrift Physics in Medicine & Biology, online erschienen am 25.03.2026. Verfasst von Didier Lustermans und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology (Maastro), GROW Research Institute for Oncology and Reproduction, Maastricht University Medical Centre+, Maastricht,…. Hauptthemen laut MEDLINE: Phantoms, Imaging, Tomography, X-Ray Computed, Photons, Radiation Dosage. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 58 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective. In proton therapy, dual-energy computed tomography (DECT) has shown to improve proton range estimation and treatment planning, but it is not yet widely implemented clinically as technology differences may restrict application (e.g. field-of-view or temporal separation). Recently, photon-counting CT (PCCT) was introduced clinically. PCCT offers multi-energy CT imaging in a single scan and with low imaging dose options. Therefore, this study aimed to evaluate proton range differences between PCCT and DECT across varying anthropomorphic phantom sizes, tissue types, and imaging doses. Approach. Virtual monoenergetic images (70 and 180 keV) were generated from DECT (80/140 kVp) and PCCT (120 kVp) scans at three imaging doses (CTDI vol,32cm : 5, 10, and 20 mGy) using a commercial head phantom and a 3D-printed abdomen phantom with exchangeable fat rings for varying dimensions. Proton therapy plans were created in RayStation (RaySearch laboratories) with varying proton beam configurations (e.g. crossing different tissues) to assess dependency of phantom size, tissue type, and proton energy. The proton range, R 80 , was extracted for each plan, and range differences were evaluated both between imaging doses and between modalities (DECT vs PCCT) at 20 mGy (inter-modality). Main results. Under reduced imaging dose, DECT showed substantial proton range differences for the large phantom sizes. It resulted in median range differences up to 20.1 mm due to visible imaging artifacts. Similar trends were seen for long proton path lengths. In contrast, for PCCT, range deviations stayed below 0.5 mm across all phantom sizes and through all tissue types. Inter-modality range variations increased when protons traversed denser tissues or longer path lengths. Significance. This work showed consistent R 80 estimations for PCCT across phantom sizes and traversing different tissue types, even under reduced imaging dose in PCCT. This offers possibilities to lower the imaging dose and thus cumulative patient dose, without compromising treatment accuracy.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingTomography, X-Ray ComputedPhotonsRadiation DosageProtonsHumansProton Therapy

Stichwörter der Autor:innen

anthropomorphic phantomslow-dose CT imagingphoton-counting CTproton range uncertaintyproton therapyradiotherapy dose calculationvirtual monoenergetic imaging

Substanzen

Protons

Autor:innen und Institutionen

  • Department of Radiation Oncology (Maastro), GROW Research Institute for Oncology and Reproduction, Maastricht University Medical Centre+, Maastricht, The Netherlands.
  • Department of Radiology and Nuclear Medicine, Maastricht University Medical Center, Maastricht, The Netherlands.
  • CARIM Cardiovascular Research Institute Maastricht, Maastricht University, Maastricht, The Netherlands.
  • Department of Radiation Oncology, LMU University Hospital, LMU Munich, Germany.
  • German Cancer Consortium (DKTK), partner site Munich, a partnership between DKFZ and LMU University Hospital Munich, Munich, Germany.
  • Bavarian Cancer Research Center (BZKF), Munich, Germany.

ORCID-Profile: Didier Lustermans, Gabriel Paiva Fonseca, Gloria Vilches-Freixas, Eva Hoeijmakers, Cécile Rlpn Jeukens, Katia Parodi, Guillaume Landry, Vicki Trier Taasti, Frank Verhaegen

Zitieren

APA

Lustermans, D., Fonseca, G. P., Vilches-Freixas, G., Rinaldi, I., Hoeijmakers, E., Jeukens, C. R., Parodi, K., Landry, G., Taasti, V. T., & Verhaegen, F. (2026). Evaluation of proton range differences in photon-counting and dual-energy computed tomography across imaging doses and anthropomorphic phantom sizes. Physics in Medicine & Biology, 71(6), 065008. https://doi.org/10.1088/1361-6560/ae50c8

Vancouver

Lustermans D, Fonseca GP, Vilches-Freixas G, Rinaldi I, Hoeijmakers E, Jeukens CR, et al. Evaluation of proton range differences in photon-counting and dual-energy computed tomography across imaging doses and anthropomorphic phantom sizes. Physics in Medicine & Biology. 2026;71(6):065008. doi:10.1088/1361-6560/ae50c8

BibTeX
@article{Lustermans2026,
  author = {Lustermans, Didier and Fonseca, Gabriel Paiva and Vilches-Freixas, Gloria and Rinaldi, Ilaria and Hoeijmakers, Eva and Jeukens, Cécile Rlpn and Parodi, Katia and Landry, Guillaume and Taasti, Vicki Trier and Verhaegen, Frank},
  title = {Evaluation of proton range differences in photon-counting and dual-energy computed tomography across imaging doses and anthropomorphic phantom sizes},
  journal = {Physics in Medicine & Biology},
  year = {2026},
  volume = {71},
  number = {6},
  pages = {065008},
  doi = {10.1088/1361-6560/ae50c8}
}
Kostenlos starten →