Neutron contribution depending on the treatment room during pediatric pencil beam scanning proton therapy

Johannes Tjelta iD, Kristian Smeland Ytre-Hauge iD, Franz Siegfried Englbrecht iD, George Dedes iD, Katia Parodi iD, Camilla Grindeland Boer iD, Helge Henjum iD, Camilla Hanquist Stokkevåg iD

Physics in Medicine & Biology · 2025

DOI: 10.1088/1361-6560/ade6be ↗

Kurz gefasst

Fachartikel in der Zeitschrift Physics in Medicine & Biology, 2025. Verfasst von Johannes Tjelta und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Oncology and Medical Physics, Haukeland University Hospital, Bergen, Norway. Hauptthemen laut MEDLINE: Proton Therapy, Neutrons. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 53 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Norwegian Childhood Cancer Society, Norges Forskningsråd, Barnekreftforeningen.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

This study compares secondary neutron fields for two proton treatment rooms to identify the impact of room geometry and treatment machine design. The equivalent dose exposure to paediatric craniospinal irradiation (CSI) patients including treatment room components was further quantified.

Methods

Treatment rooms in Munich and Bergen were implemented in FLUKA. 75 and 200 MeV proton beams were simulated to a PMMA phantom to investigate scoring regions at 2 m from the isocenter at 0°, 45°, 90° and 135°, with respect to the beam. Neutron spectra were collected for individual room components. Ambient dose equivalent tables were used to convert neutron fluence to dose and compare the rooms. Paediatric CSI treatment plans were simulated within the Bergen facility, and the equivalent dose to the patients was cataloged according to the component of the final neutron interaction.

Results

The phantom simulations had similar total ambient dose equivalent, with values of (0.8–1.5) μ Sv/Gy for the 75 MeV beam and (4–25) μ Sv/Gy for the 200 MeV beam in scoring volumes at 0°–135°. The dose contribution from the concrete room in Bergen exceeded the Munich room (0.07 ± 0.01 versus 0.02 ± 0.01 μ Sv/Gy) whereas the contribution from the Munich gantry was higher (0.020 ± 0.003 versus 0.005 ± 0.001 μ Sv/Gy). The resulting overall neutron equivalent dose to the organs ranged (0.09–2.96) mSv/Gy for the CSI plans and (0.004–0.05) mSv/Gy for the boost plans. In certain scenarios, the patient frame, gantry and concrete room contributed up to 44% to the overall neutron equivalent dose for organs far away from the treatment fields. Significance. Inclusion of the treatment room increased the accuracy of the calculated dose (with a difference of up to 0.16 mSv/Gy) associated with personalized proton therapy, especially in organs distant from the treatment field. The relatively small differences in neutron ambient dose equivalent levels found across the two facilities suggests that the presented data can be applied where detailed room simulations are not available.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Proton TherapyNeutronsHumansPhantoms, ImagingChildRadiotherapy DosageRadiotherapy Planning, Computer-AssistedCraniospinal Irradiation

Stichwörter der Autor:innen

FLUKAMonte Carlo modellingequivalent doseneutron spectroscopypaediatric patientproton therapy

Autor:innen und Institutionen

  • Department of Oncology and Medical Physics, Haukeland University Hospital, Bergen, Norway.
  • Department of Physics and Technology, University of Bergen, Bergen, Norway.
  • Department of Medical Physics, Ludwig-Maximilians-Universität München (LMU Munich), Garching b, Munich, Germany.

ORCID-Profile: Johannes Tjelta, Kristian Smeland Ytre-Hauge, Franz Siegfried Englbrecht, George Dedes, Katia Parodi, Camilla Grindeland Boer, Helge Henjum, Camilla Hanquist Stokkevåg

Förderung

  • Norwegian Childhood Cancer Society
  • Norges Forskningsråd
  • Barnekreftforeningen

Zitieren

APA

Tjelta, J., Ytre-Hauge, K. S., Englbrecht, F. S., Dedes, G., Parodi, K., Boer, C. G., Henjum, H., & Stokkevåg, C. H. (2025). Neutron contribution depending on the treatment room during pediatric pencil beam scanning proton therapy. Physics in Medicine & Biology. https://doi.org/10.1088/1361-6560/ade6be

Vancouver

Tjelta J, Ytre-Hauge KS, Englbrecht FS, Dedes G, Parodi K, Boer CG, et al. Neutron contribution depending on the treatment room during pediatric pencil beam scanning proton therapy. Physics in Medicine & Biology. 2025. doi:10.1088/1361-6560/ade6be

BibTeX
@article{Tjelta2025,
  author = {Tjelta, Johannes and Ytre-Hauge, Kristian Smeland and Englbrecht, Franz Siegfried and Dedes, George and Parodi, Katia and Boer, Camilla Grindeland and Henjum, Helge and Stokkevåg, Camilla Hanquist},
  title = {Neutron contribution depending on the treatment room during pediatric pencil beam scanning proton therapy},
  journal = {Physics in Medicine & Biology},
  year = {2025},
  doi = {10.1088/1361-6560/ade6be}
}
Kostenlos starten →