Development of a breathing lung phantom for proton CT imaging
Physics in Medicine & Biology · 2025
DOI: 10.1088/1361-6560/ade2b6 ↗Kurz gefasst
Fachartikel in der Zeitschrift Physics in Medicine & Biology, 2025. Verfasst von V Wegener und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Physics, Ludwig-Maximilians-Universität München (LMU), Munich, Germany. Hauptthemen laut MEDLINE: Phantoms, Imaging, Tomography, X-Ray Computed, Respiration, Lung. Die Arbeit nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Deutsche Forschungsgemeinschaft—DFG, Deutsche Forschungsgemeinschaft.
Abstract (Originalsprache)
Objective. To report on the design of a deformable lung phantom capable of imitating breathing motion with realistic tissue surrogate properties for proton imaging applications. Approach . The phantom was manufactured via 3D printing and silicone moulding, with a customised structural design for motor-controlled breathing motion. The overall size of the phantom was rescaled to fit in the experimental proton computed tomography (pCT) scanner prototype, featuring a 284 mm maximum size for the imaging field-of-view. Several flexible resins were evaluated in perspective of flexibility by varying ultraviolet exposure times, as increased exposure results in resin hardening at each layer. We optimised the structure to achieve ideal lung compression properties, while preserving its integrity to hold the weight of a solid tumour. Phantom material properties were characterised by segmentation of each component in x-ray CT and pCT images, to determine the CT number expressed in Hounsfield units and the relative stopping power (RSP) with respect to water. Main results . We achieved non-homogenous compression in the lung using a grid structure with gradient thickness. The rigid ribcage was 3D printed using granite based material. The tumour motion implemented in the phantom design, as measured using template-matching in fluoroscopic x-ray imaging, revealed hysteretic motion with 10 mm peak-to-peak in the superior–inferior direction. Significance. The developed deformable lung phantom imitated lung motion characteristics, featuring CT number and RSP values in the range comparable to human tissues. The developed breathing phantom is put forward for experimental motion studies in pCT imaging.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Medical Physics, Ludwig-Maximilians-Universität München (LMU), Munich, Germany.
- Department of Radiation Oncology, LMU University Hospital, Ludwig-Maximilians-Universität München, Munich, Germany.
- German Cancer Consortium (DKTK), partner site Munich, a partnership between DKFZ and LMU University Hospital, Munich, Germany.
- Bavarian Cancer Research Center (BZKF), Munich, Germany.
- Division of Biomedical Engineering Sciences, Loma Linda University, Loma Linda, CA, United States of America.
- Northwestern Medicine Chicago Proton Center, Warrenville, IL, United States of America.
ORCID-Profile: V Wegener, M Rabe, G Landry, R W Schulte, K Parodi, J Bortfeld, R P Johnson, J Lascaud, M Riboldi, P Palaniappan
Förderung
- Deutsche Forschungsgemeinschaft—DFG
- Deutsche Forschungsgemeinschaft
Zitieren
APA
Wegener, V., Fischer, T., Rabe, M., Landry, G., Schulte, R. W., Parodi, K., Bortfeld, J., Pankuch, M., Johnson, R. P., Lascaud, J., Dedes, G., Riboldi, M., & Palaniappan, P. (2025). Development of a breathing lung phantom for proton CT imaging. Physics in Medicine & Biology. https://doi.org/10.1088/1361-6560/ade2b6
Vancouver
Wegener V, Fischer T, Rabe M, Landry G, Schulte RW, Parodi K, et al. Development of a breathing lung phantom for proton CT imaging. Physics in Medicine & Biology. 2025. doi:10.1088/1361-6560/ade2b6
BibTeX
@article{Wegener2025,
author = {Wegener, V and Fischer, T and Rabe, M and Landry, G and Schulte, R W and Parodi, K and Bortfeld, J and Pankuch, M and Johnson, R P and Lascaud, J and Dedes, G and Riboldi, M and Palaniappan, P},
title = {Development of a breathing lung phantom for proton CT imaging},
journal = {Physics in Medicine & Biology},
year = {2025},
doi = {10.1088/1361-6560/ade2b6}
}
EUVIMED