Bubble cloud characteristics and ablation efficiency in dual-frequency intrinsic threshold histotripsy
Physics in Medicine & Biology · 2023
DOI: 10.1088/1361-6560/ad00a5 ↗Kurz gefasst
Fachartikel in der Zeitschrift Physics in Medicine & Biology, 2023. Verfasst von Connor Edsall und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering and Mechanics, Virginia Polytechnic Institute and State University, 325 Stanger St., Blacksburg, VA, 24061, Unit…. Hauptthemen laut MEDLINE: High-Intensity Focused Ultrasound Ablation, Lithotripsy. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCI NIH HHS, National Institutes of Health.
Abstract (Originalsprache)
Histotripsy is a non-thermal focused ultrasound ablation method that destroys tissue through the generation and activity of acoustic cavitation bubble clouds. Intrinsic threshold histotripsy uses single-cycle pulses to generate bubble clouds when the dominant negative pressure phase exceeds an intrinsic threshold of ∼25–30 MPa. The ablation efficiency is dependent upon the size and density of bubbles within the bubble cloud. This work investigates the effects of dual-frequency pulsing schemes on the bubble cloud behavior and ablation efficiency in intrinsic threshold histotripsy. A modular 500 kHz:3 MHz histotripsy transducer treated agarose phantoms using dual-frequency histotripsy pulses with a 1:1 pressure ratio from 500 kHz and 3 MHz frequency elements and varying arrival times for the 3 MHz pulse relative to the arrival of the 500 kHz pulse (−100 ns, 0 ns, and +100 ns). High-speed optical imaging captured cavitation effects to characterize bubble cloud and individual bubble dynamics. The effects of dual-frequency pulsing on lesion formation and ablation efficiency were also investigated in red blood cell (RBC) phantoms. Results showed that the single bubble and bubble cloud size for dual-frequency cases were intermediate to published results for the component single-frequencies of 500 kHz and 3 MHz. Additionally, bubble cloud size and dynamics were shown to be altered by the arrival time of the 3 MHz pulse with respect to the 500 kHz pulse, with more uniform cloud expansion and collapse observed for early (−100 ns) arrival. Finally, RBC phantom experiments showed that dual-frequency exposures were capable of generating precise lesions with smaller areas and higher ablation efficiencies than previously published results for 500 kHz or 3 MHz. Overall, results demonstrate dual-frequency histotripsy’s ability to modulate bubble cloud size and dynamics can be leveraged to produce precise lesions at higher ablation efficiencies than previously observed for single-frequency pulsing.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Biomedical Engineering and Mechanics, Virginia Polytechnic Institute and State University, 325 Stanger St., Blacksburg, VA, 24061, United States of America.
- Department of Materials Science and Engineering, Virginia Polytechnic Institute and State University, 445 Old Turner St., Blacksburg, VA 24061, United States of America.
- Biomedical Engineering, University of Michigan, Carl A. Gerstacker Building, 2200 Bonisteel Blvd, Ann Arbor, MI 48109-2133, United States of America.
- ICTAS Center for Engineered Health, Virginia Polytechnic Institute and State University, 325 Stanger St., Blacksburg, VA, 24061, United States of America.
ORCID-Profile: Connor Edsall, Timothy L Hall, Eli Vlaisavljevich
Förderung
- NCI NIH HHS
- National Institutes of Health
Zitieren
APA
Edsall, C., Huynh, L., Hall, T. L., & Vlaisavljevich, E. (2023). Bubble cloud characteristics and ablation efficiency in dual-frequency intrinsic threshold histotripsy. Physics in Medicine & Biology. https://doi.org/10.1088/1361-6560/ad00a5
Vancouver
Edsall C, Huynh L, Hall TL, Vlaisavljevich E. Bubble cloud characteristics and ablation efficiency in dual-frequency intrinsic threshold histotripsy. Physics in Medicine & Biology. 2023. doi:10.1088/1361-6560/ad00a5
BibTeX
@article{Edsall2023,
author = {Edsall, Connor and Huynh, Laura and Hall, Timothy L and Vlaisavljevich, Eli},
title = {Bubble cloud characteristics and ablation efficiency in dual-frequency intrinsic threshold histotripsy},
journal = {Physics in Medicine & Biology},
year = {2023},
doi = {10.1088/1361-6560/ad00a5}
}
EUVIMED