Fachartikel in der Zeitschrift Nanotechnology, online erschienen am 21.08.2023. Verfasst von Baoqing Han und 10 weiteren Autor:innen. Beteiligte Einrichtung: School of Mechano- Electronic Engineering, Xidian University, Xi'an 710071, People's Republic of China. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 44 Literaturangaben. Gefördert von: Natural Science Basic Research Program of Shaanxi, National Natural Science Foundation of China, Fundamental Research Funds for the Central Universities.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Isoprene is a typical physiological marker that can be used to screen for chronic liver disease. This work developed a portable micro-integrated chromatography analysis system based on micro-electromechanical system technology, nanomaterials technology and embedded microcontroller technology. The system integrated components such as graphene oxide quantum dots modified semi-packed microcolumn, In 2 O 3 nanoflower (NF) gas-sensitive detector and 3D printed miniature solenoid valve group. The effectiveness of the separation effect of the micro-integrated system was verified by gas mixture test; the laws of the influence of carrier gas pressure and column temperature on the chromatographic separation performance, respectively, were investigated, and the working conditions (column temperature 90 °C and carrier gas pressure 7.5 kPa) for system testing were determined. The percentages of relative standard deviation of the peak areas and retention times obtained for the separated gases were in the range of 0.95%–6.06%, indicating the good reproducibility of the system. Meanwhile, the microintegrated system could detect isoprene down to 50 ppb at small injection volume (1 ml). The system response increased with increasing isoprene concentration and was linearly correlated with isoprene concentration ( R 2 = 0.986), indicating that the system was expected to be used for trace detection of isoprene, a marker gas for liver disease, in the future.

Schlagwörter

Stichwörter der Autor:innen

isoprenemicro columnmicro gas chromatography systemmicro gas sensor

Autor:innen und Institutionen

  • School of Mechano- Electronic Engineering, Xidian University, Xi'an 710071, People's Republic of China.
  • State Key Laboratory for Manufacturing Systems Engineering, Xi'an Jiaotong University, Xi'an, Shaanxi 710049, People's Republic of China.
  • School of Information and Control Engineering, Xi'an University of Architecture and Technology, Xi'an, 710055, People's Republic of China.
  • Guangzhou Institute of Technology, Xidian University, Guangzhou 510555, People's Republic of China.
  • Quzhou Peoples Hosp, Dept Neurol, 2 Zhongloudi Rd, Quzhou 324000, People's Republic of China.

ORCID-Profile: Baoqing Han

Förderung

  • Natural Science Basic Research Program of Shaanxi
  • National Natural Science Foundation of China
  • Fundamental Research Funds for the Central Universities
  • National Defense Basic Scientific Research Program of China
  • Youth Innovation Team of Shaanxi Universities
  • Wuhu and Xidian University Special Fund for Industry-University-Research Cooperation

Zitieren

APA

Han, B., Zhang, X., Wang, Y., Wang, W., Wang, B., Li, S., Wang, H., Yan, Y., Han, J., Wang, C., & Wang, C. (2023). Real-time detection of isoprene marker gas based on micro-integrated chromatography system with GOQDs-modified μ GC column and metal oxide gas detector. Nanotechnology, 34(45), 455501. https://doi.org/10.1088/1361-6528/aced10

Vancouver

Han B, Zhang X, Wang Y, Wang W, Wang B, Li S, et al. Real-time detection of isoprene marker gas based on micro-integrated chromatography system with GOQDs-modified μ GC column and metal oxide gas detector. Nanotechnology. 2023;34(45):455501. doi:10.1088/1361-6528/aced10

BibTeX
@article{Han2023,
  author = {Han, Baoqing and Zhang, Xinyu and Wang, Yan and Wang, Wenjuan and Wang, Benben and Li, Shuai and Wang, Hairong and Yan, Yuefei and Han, Jiusheng and Wang, Chuanliu and Wang, Congsi},
  title = {Real-time detection of isoprene marker gas based on micro-integrated chromatography system with GOQDs-modified μ GC column and metal oxide gas detector},
  journal = {Nanotechnology},
  year = {2023},
  volume = {34},
  number = {45},
  pages = {455501},
  doi = {10.1088/1361-6528/aced10}
}
Kostenlos starten →