Physics-informed neural-network calibration of pump–probe nonorthogonality in SERF co-magnetometers
Measurement Science and Technology · 14.08.2026 · 37 (32) · 325902
DOI: 10.1088/1361-6501/ae9330 ↗Kurz gefasst
Fachartikel in der Zeitschrift Measurement Science and Technology, online erschienen am 12.08.2026. Verfasst von Jiahang Li und 4 weiteren Autor:innen. Die Arbeit nennt 37 Literaturangaben. Gefördert von: Beijing Natural Science Foundation, Natural Science Foundation of Zhejiang Province.
Abstract (Originalsprache)
Pump–probe nonorthogonality causes readout bias in spin-exchange relaxation-free (SERF) co-magnetometers through longitudinal leakage into the Faraday detection channel. This paper develops a pump-power-scan calibration method and a physics-informed neural network (PINN) inverse estimator for determining the pump–probe nonorthogonality angle. The method exploits pump-power-induced changes in the longitudinal polarization, transverse response, and steady readout signal. The actual pump beam defines the experimental z -axis, and the actual probe direction is represented in this pump-fixed frame by an out-of-plane nonorthogonality angle and an in-plane skew angle. The proposed PINN embeds the pump-power-dependent optical response and the measured readout into the inverse estimation loss, enabling angle recovery from experimentally recorded scans with independent transverse-response information. An experimental dual-axis scan with calibrated detection gain recovers both probe-direction angles, with the estimates reported at a conservative precision of 0.05 mrad. A second scan covering the 14-point pump-power grid returns out-of-plane and in-plane skew estimates of 1.35 mrad and − 1.65 mrad, consistent with the reference settings within a ± 0.05 mrad repeatability band. Transverse-response checks are consistent with the sign reversal of the in-plane skew pathway and the stability of the two recovered angles. A linearized effective pump-power fit on the same 14-point grid achieves an RMSE of 0.0295 a.u. and a coefficient of determination of 0.9981. These results show that pump-power modulation provides an experimentally accessible response path for PINN-based pump–probe nonorthogonality calibration in SERF co-magnetometers.
Förderung
- Beijing Natural Science Foundation
- Natural Science Foundation of Zhejiang Province
Zitieren
APA
Li, J., Wang, Z., Li, F., Wang, R., & Pang, H. (2026). Physics-informed neural-network calibration of pump–probe nonorthogonality in SERF co-magnetometers. Measurement Science and Technology, 37(32), 325902. https://doi.org/10.1088/1361-6501/ae9330
Vancouver
Li J, Wang Z, Li F, Wang R, Pang H. Physics-informed neural-network calibration of pump–probe nonorthogonality in SERF co-magnetometers. Measurement Science and Technology. 2026;37(32):325902. doi:10.1088/1361-6501/ae9330
BibTeX
@article{Li2026,
author = {Li, Jiahang and Wang, Zhuo and Li, Feng and Wang, Ruigang and Pang, Haoying},
title = {Physics-informed neural-network calibration of pump–probe nonorthogonality in SERF co-magnetometers},
journal = {Measurement Science and Technology},
year = {2026},
volume = {37},
number = {32},
pages = {325902},
doi = {10.1088/1361-6501/ae9330}
}
EUVIMED