EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

AFM z -piezo calibration method for height measurements in the sub-nanometer to nanometer range using stepped surfaces

Gheorghe Stan

Measurement Science and Technology · 21.08.2026 · 37 (33) · 335010

DOI: 10.1088/1361-6501/ae8568 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Measurement Science and Technology, online erschienen am 19.08.2026. Verfasst von Gheorghe Stan. Die Arbeit nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Institute of Standards and Technology, U.S. Department of Commerce, CHIPS for America.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

In this work, we present a method for calibrating height measurements from the z -piezo of an atomic force microscope (AFM) using stepped surfaces with subnanometer staircases. The method analyzes common trace and retrace images obtained from AFM topography scans across straight terraces of these stepped surfaces. By combining these data from a single scan, we can determine both the quantitative height calibration and the linearity of the AFM z -piezo across a full range of heights. This eliminates the need to separately image artifacts of different heights. This approach is significant because it incorporates all measured step heights from a single scan, allowing calibration for any height difference between them. As a result, it serves as a multipoint calibration method that spans a height range from subnanometer levels to tens of nanometers. This contrasts with single-point methods, which typically calibrate against a single height derived from measurements of a single artifact. Our calibration method relies on straightforward metrics, including average measured step heights, standard errors from the mean, relative accuracy, and percentage uncertainty. The most sensitive metrics for detecting potential miscalibration and nonlinearity in the AFM z -piezo are the average residuals and relative accuracy. In contrast, the standard errors and percentage uncertainty primarily relate to the roughness of the stepped surface. We demonstrate the application of this calibration method on scans over a 6H-SiC(0001) stepped surface, with calibrated heights ranging from a minimum step height of 0.756 nm to heights approaching 50 nm.

Förderung

  • National Institute of Standards and Technology, U.S. Department of Commerce
  • CHIPS for America

Zitieren

APA

Stan, G. (2026). AFM z -piezo calibration method for height measurements in the sub-nanometer to nanometer range using stepped surfaces. Measurement Science and Technology, 37(33), 335010. https://doi.org/10.1088/1361-6501/ae8568

Vancouver

Stan G. AFM z -piezo calibration method for height measurements in the sub-nanometer to nanometer range using stepped surfaces. Measurement Science and Technology. 2026;37(33):335010. doi:10.1088/1361-6501/ae8568

BibTeX
@article{Stan2026,
  author = {Stan, Gheorghe},
  title = {AFM z -piezo calibration method for height measurements in the sub-nanometer to nanometer range using stepped surfaces},
  journal = {Measurement Science and Technology},
  year = {2026},
  volume = {37},
  number = {33},
  pages = {335010},
  doi = {10.1088/1361-6501/ae8568}
}
Kostenlos starten →