Interferometric laser-beam characterisation with computed spherical reference wave
Measurement Science and Technology · 19.06.2026 · 37 (25) · 255202
DOI: 10.1088/1361-6501/ae7aa9 ↗Kurz gefasst
Fachartikel in der Zeitschrift Measurement Science and Technology, online erschienen am 23.06.2026. Verfasst von B Andreas. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 45 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
For interferometric length, velocity, or acceleration measurements with challenging uncertainty requirements, a contribution which is caused by diffraction, needs to be considered. To assess this contribution it is necessary to characterise the laser beams used for respective interferometers. Here, I review the shortcomings of the commonly used beam characterisation methods and present the proof of principle of an alternative method which utilises interference with a computed reference wave to recover the complex amplitude of a sample beam with at least one order of magnitude increased sensitivity in the low-irradiance tail-regions of a laser beam. It is also shown how the relative length error of a sample beam is obtained from the complex amplitude.
Zitieren
APA
Andreas, B. (2026). Interferometric laser-beam characterisation with computed spherical reference wave. Measurement Science and Technology, 37(25), 255202. https://doi.org/10.1088/1361-6501/ae7aa9
Vancouver
Andreas B. Interferometric laser-beam characterisation with computed spherical reference wave. Measurement Science and Technology. 2026;37(25):255202. doi:10.1088/1361-6501/ae7aa9
BibTeX
@article{Andreas2026,
author = {Andreas, B},
title = {Interferometric laser-beam characterisation with computed spherical reference wave},
journal = {Measurement Science and Technology},
year = {2026},
volume = {37},
number = {25},
pages = {255202},
doi = {10.1088/1361-6501/ae7aa9}
}
EUVIMED