Noncontact MEMS thermal flow sensors integrated in glass microfluidic chemical chip
Journal of Micromechanics and Microengineering · 30.06.2025 · 35 (6) · 065013
DOI: 10.1088/1361-6439/ade162 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Micromechanics and Microengineering, online erschienen am 23.06.2025. Verfasst von Shao-Yang Hung und 6 weiteren Autor:innen. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 46 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Science and Technology Council.
Abstract (Originalsprache)
This research focuses on developing an integrated flow rate sensor for microfluidic systems, widely used in medical and chemical applications. Unlike commercial sensors that should be attached externally, this sensor is embedded directly on the chip, saving space and improving low-flow measurement accuracy. For chemical experiment use, glass substrates were used for the microfluidic channel. In addition, by isolating metal components from fluid exposure, the design avoids contamination and unwanted electrochemical reactions in the liquid flow phase during the flow rate sensing. The sensor operates on a calorimetric thermal flow-sensing principle, allowing flexible placement across the chip. Using multi-physics simulations, optimal sensor geometry and dimension were determined, and microfabrication processes like photolithography and metal deposition were employed to realize the chip. Testing shows it can operate at heating levels of up to 75 °C and measure low flow rates from 0 to 8 μ l min −1 with linear sensitivity.
Förderung
- National Science and Technology Council
Zitieren
APA
Hung, S., Lin, Z., Fujita, H., Li, S., Chen, C., Kitamori, T., & Morikawa, K. (2025). Noncontact MEMS thermal flow sensors integrated in glass microfluidic chemical chip. Journal of Micromechanics and Microengineering, 35(6), 065013. https://doi.org/10.1088/1361-6439/ade162
Vancouver
Hung S, Lin Z, Fujita H, Li S, Chen C, Kitamori T, et al. Noncontact MEMS thermal flow sensors integrated in glass microfluidic chemical chip. Journal of Micromechanics and Microengineering. 2025;35(6):065013. doi:10.1088/1361-6439/ade162
BibTeX
@article{Hung2025,
author = {Hung, Shao-Yang and Lin, Zhong-Wei and Fujita, Hiroyuki and Li, Sheng-Shian and Chen, Chihchen and Kitamori, Takehiko and Morikawa, Kyojiro},
title = {Noncontact MEMS thermal flow sensors integrated in glass microfluidic chemical chip},
journal = {Journal of Micromechanics and Microengineering},
year = {2025},
volume = {35},
number = {6},
pages = {065013},
doi = {10.1088/1361-6439/ade162}
}