A pre-clinical application study of the SATA-LRS laparoscopic instrument in a human cadaver model

Tomas Lenssen, Roelf R Postema iD, Christian Camenzuli, Jean Calleja-Agius, Jenny Dankelman, Tim Horeman-Franse

Minimally Invasive Therapy & Allied Technologies · 02.11.2025 · 34 (6) · 447-455

DOI: 10.1080/13645706.2025.2554077 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Minimally Invasive Therapy & Allied Technologies, online erschienen am 18.09.2025. Verfasst von Tomas Lenssen und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomechanical Engineering, Delft University of Technology, Delft, The Netherlands. Hauptthemen laut MEDLINE: Laparoscopy, Hernia, Inguinal, Herniorrhaphy. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 15 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: ZonMW through the Project SMART Surgical System: High-Quality Medical Devices.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: To validate whether the SATA-LRS, a novel reusable articulating laparoscopic instrument, fits surgical practice, a pre-clinical study was performed. METHODS: Thirteen medical doctors used the instrument in a laparoscopic endoscopic inguinal hernia repair (TEP)-like task inside a cadaver. A set of sensors on the instrument handle detected motion and articulation of the instrument tip. Data from the sensors and video recordings were used to assess the amount and type of movement of the instrument and the time spent on tasks. questionnaire was used to gain insight into the participants' perception of the contextual factors. RESULTS: There was no difference between task time and instrument tip velocities when using articulation (or not) and all participants used articulation at least half of the task time. Instrument-handle movement, indicating the user's hand and arm movement, was significantly reduced when using articulation. The questionnaire indicated strong acceptance of the instrument and the experimental setup, and a desire to use the instrument in surgery by most participants. CONCLUSIONS: The added articulation feature of the SATA-LRS instrument was deemed beneficial by the participants, showed no increased handling complexity or time spent on the task and was used frequently when enabled, indicating intuitiveness.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

LaparoscopyHernia, InguinalHerniorrhaphyHumansCadaverEquipment DesignSurveys and QuestionnairesMaleFemaleVideo Recording

Stichwörter der Autor:innen

IMUSATAclinicallaparoscopyvalidation

Autor:innen und Institutionen

  • Department of Biomechanical Engineering, Delft University of Technology, Delft, The Netherlands.
  • Department of Surgery, University Medical Centers Amsterdam, Amsterdam, The Netherlands.
  • Department of Surgery, University of Malta, L-Università ta' Malta, Msida, Malta.
  • Department of Anatomy, Faculty of Medicine and Surgery, University of Malta, L-Università ta' Malta, Msida, Malta.

ORCID-Profile: Roelf R Postema

Förderung

  • ZonMW through the Project SMART Surgical System: High-Quality Medical Devices

Zitieren

APA

Lenssen, T., Postema, R. R., Camenzuli, C., Calleja-Agius, J., Dankelman, J., & Horeman-Franse, T. (2025). A pre-clinical application study of the SATA-LRS laparoscopic instrument in a human cadaver model. Minimally Invasive Therapy & Allied Technologies, 34(6), 447-455. https://doi.org/10.1080/13645706.2025.2554077

Vancouver

Lenssen T, Postema RR, Camenzuli C, Calleja-Agius J, Dankelman J, Horeman-Franse T. A pre-clinical application study of the SATA-LRS laparoscopic instrument in a human cadaver model. Minimally Invasive Therapy & Allied Technologies. 2025;34(6):447-455. doi:10.1080/13645706.2025.2554077

BibTeX
@article{Lenssen2025,
  author = {Lenssen, Tomas and Postema, Roelf R and Camenzuli, Christian and Calleja-Agius, Jean and Dankelman, Jenny and Horeman-Franse, Tim},
  title = {A pre-clinical application study of the SATA-LRS laparoscopic instrument in a human cadaver model},
  journal = {Minimally Invasive Therapy & Allied Technologies},
  year = {2025},
  volume = {34},
  number = {6},
  pages = {447--455},
  doi = {10.1080/13645706.2025.2554077}
}
Kostenlos starten →