Wireless temperature measurement based on magnetic harmonic signal during magnetic heating operation for hyperthermia
AIP Advances · 01.02.2026 · 16 (2)
DOI: 10.1063/9.0001070 ↗Kurz gefasst
Fachartikel in der Zeitschrift AIP Advances, online erschienen am 12.02.2026. Verfasst von Akihiro Kuwahata und 3 weiteren Autor:innen. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Nakatani Foundation for Advancement of Measuring Technologies in Biomedical Engineering, Iketani Science and Technology Foundation, Japan Agency for Medical Research and Development.
Abstract (Originalsprache)
We developed a wireless temperature measurement method based on magnetic harmonic signals during magnetic heating and evaluated highly concentrated magnetic nanoparticles (MNPs) for hyperthermia applications. XRD confirmed that the iron oxide composition of Resovist® remained unchanged after concentration, while TEM and DLS analyses demonstrated reversible dispersibility between concentrated and diluted states. Magnetization measurements showed an increase in saturation magnetization, and the AC hysteresis loop area of fourfold-concentrated Resovist® was 3.8 times larger than that of the original sample, indicating enhanced heating efficiency. The detectable distance and volume for temperature measurement improved from 5 mm to 12 mm and from 10 µL to 5 µL, respectively. The developed system achieved noninvasive temperature measurement with an error of 0.07°C and constant-temperature control with a deviation of 0.56°C, demonstrating sufficient precision for biomedical applications.
Förderung
- Nakatani Foundation for Advancement of Measuring Technologies in Biomedical Engineering
- Iketani Science and Technology Foundation
- Japan Agency for Medical Research and Development
- Tohoku University Business Incubation Program
- JSPS KAKENHI
- JST FOREST Program
Zitieren
APA
Kuwahata, A., Yamazaki, A., Shinohara, R., & Yabukami, S. (2026). Wireless temperature measurement based on magnetic harmonic signal during magnetic heating operation for hyperthermia. AIP Advances, 16(2). https://doi.org/10.1063/9.0001070
Vancouver
Kuwahata A, Yamazaki A, Shinohara R, Yabukami S. Wireless temperature measurement based on magnetic harmonic signal during magnetic heating operation for hyperthermia. AIP Advances. 2026;16(2). doi:10.1063/9.0001070
BibTeX
@article{Kuwahata2026,
author = {Kuwahata, Akihiro and Yamazaki, Aoi and Shinohara, Riku and Yabukami, Shin},
title = {Wireless temperature measurement based on magnetic harmonic signal during magnetic heating operation for hyperthermia},
journal = {AIP Advances},
year = {2026},
volume = {16},
number = {2},
doi = {10.1063/9.0001070}
}