In-line HCl/NaClO decontamination for carryover control in piezoelectric droplet-printing digital PCR

Zhaonan Liu, Hongwei Quan, Mengqi Kong, Zonghao Li, Yuntong Gao, Deding Tang, Yan Gong, Jie Hu

Biomicrofluidics · 01.12.2026 · 20 (6)

DOI: 10.1063/5.0358108 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biomicrofluidics, online erschienen am 17.11.2026. Verfasst von Zhaonan Liu und 7 weiteren Autor:innen. Die Arbeit nennt 30 Literaturangaben. Gefördert von: Scientific Research Foundation of Education Department of Anhui Province of China.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Piezoelectric droplet-on-demand printing enables controlled partition generation for digital polymerase chain reaction (dPCR) and spatially addressable droplet arrays. However, a shared dispensing nozzle introduces a risk of sample-to-sample carryover. Here, we evaluated four nozzle-cleaning strategies under fixed printing conditions: a de-ionized-water rinse, two commercial nucleic acid decontamination reagents (NH9901 and VNR11-01), and an automated in-line HCl/NaClO tip-cleaning procedure (G4). Cleaning efficacy was assessed by qPCR for carryover and inhibition and independently cross-validated by direct dPCR partition counting using two sample types. Under single-cycle conditions, G4 showed no detectable carryover: no-template control (NTC) amplification within 45 qPCR cycles and no positive dPCR partitions, without measurable inhibition in the corresponding positive controls. Zero positives among N = 19 469–20 412 analyzed NTC partitions correspond to a one-sided 95% upper bound of ≈3/N (≈1.5 × 10−4 per partition); these are reported as below the detection limit. Droplet-generation characteristics were preserved after a single cleaning cycle, and the digital signal response was proportional (FAM R2 = 0.9998; ROX R2 = 0.9972) with 3.31% within-session repeatability (n = 20). A 20-cycle assessment at a nominal 3.33 × 102 copies/μL kept NTC concentrations below the detection limit at all evaluated checkpoints with positive control (PC) concentrations of 3.62 × 102–3.64 × 102 copies/μL. Exploratory analysis of 493 clinical specimens assessed workflow applicability in complex biological matrices without diagnostic performance evaluation. This in-line decontamination strategy provides a practical approach for contamination-controlled nozzle operation in piezoelectric droplet-printing dPCR under the evaluated single-nozzle and single-platform conditions; broader robustness and long-term material compatibility remain to be established.

Förderung

  • Scientific Research Foundation of Education Department of Anhui Province of China

Zitieren

APA

Liu, Z., Quan, H., Kong, M., Li, Z., Gao, Y., Tang, D., Gong, Y., & Hu, J. (2026). In-line HCl/NaClO decontamination for carryover control in piezoelectric droplet-printing digital PCR. Biomicrofluidics, 20(6). https://doi.org/10.1063/5.0358108

Vancouver

Liu Z, Quan H, Kong M, Li Z, Gao Y, Tang D, et al. In-line HCl/NaClO decontamination for carryover control in piezoelectric droplet-printing digital PCR. Biomicrofluidics. 2026;20(6). doi:10.1063/5.0358108

BibTeX
@article{Liu2026,
  author = {Liu, Zhaonan and Quan, Hongwei and Kong, Mengqi and Li, Zonghao and Gao, Yuntong and Tang, Deding and Gong, Yan and Hu, Jie},
  title = {In-line HCl/NaClO decontamination for carryover control in piezoelectric droplet-printing digital PCR},
  journal = {Biomicrofluidics},
  year = {2026},
  volume = {20},
  number = {6},
  doi = {10.1063/5.0358108}
}
Kostenlos starten →