Time-resolved photon energy spectrum measurement of pulsed x rays from low-impedance rod-pinch diodes
Review of Scientific Instruments · 01.11.2026 · 97 (11)
DOI: 10.1063/5.0351726 ↗Kurz gefasst
Fachartikel in der Zeitschrift Review of Scientific Instruments, online erschienen am 23.11.2026. Verfasst von Liuhuo Wang und 8 weiteren Autor:innen. Beteiligte Einrichtung: Guangdong Power Grid Corporation, Guangzhou 510620, China. Die Arbeit nennt 32 Literaturangaben. Gefördert von: Guangdong Power Grid Company.
Abstract (Originalsprache)
The photon energy spectrum of pulsed x rays is an important input for flash radiographic source characterization, dose assessment, and radiation transport model validation. However, time-resolved spectrum measurement within a single nanosecond pulse remains challenging because of the high photon fluence, broad spectral distribution, and strong electromagnetic interference. To obtain the spectral evolution within a single pulse, a 16-channel silicon Si-PIN absorption-array spectrometer is developed and a corresponding time-resolved spectrum reconstruction method is proposed. To improve the numerical stability of the absorption-based measurement model, a simulated annealing algorithm is used to optimize the filter combination by minimizing the condition number of the response matrix. The response matrix is further corrected using Geant4 simulations. To address the ill-posed spectrum unfolding problem with a limited number of channels, a multilayer perceptron neural network is trained and compared with conventional iterative methods and a genetic algorithm. In simulation tests, the neural-network method achieves a root-mean-square error of 0.0036 and provides better reconstruction accuracy and robustness than other methods. In the experiments on the WRPD-1 facility, the time-integrated photon energy spectrum of WRPD-1 is reconstructed and the pulsed x-ray energy spectra within five consecutive time windows are acquired. The reconstructed spectra reveal a clear soft–hard–soft evolution during the radiation pulse, demonstrating the feasibility of the proposed method for time-resolved diagnostics of pulsed x-ray sources.
Autor:innen und Institutionen
- Guangdong Power Grid Corporation, Guangzhou 510620, China.
- Institute of Science and Education Development, Xi'an Jiaotong University, Xi'an, China.
ORCID-Profile: Jingrui Han, Tongquan Wang, Huantong Shi, Xingwen Li
Förderung
- Guangdong Power Grid Company
Zitieren
APA
Wang, L., Han, J., Wang, T., Huang, T., Diao, H., Qi, H., Ye, M., Shi, H., & Li, X. (2026). Time-resolved photon energy spectrum measurement of pulsed x rays from low-impedance rod-pinch diodes. Review of Scientific Instruments, 97(11). https://doi.org/10.1063/5.0351726
Vancouver
Wang L, Han J, Wang T, Huang T, Diao H, Qi H, et al. Time-resolved photon energy spectrum measurement of pulsed x rays from low-impedance rod-pinch diodes. Review of Scientific Instruments. 2026;97(11). doi:10.1063/5.0351726
BibTeX
@article{Wang2026,
author = {Wang, Liuhuo and Han, Jingrui and Wang, Tongquan and Huang, Tao and Diao, Haoran and Qi, Han and Ye, Mingjie and Shi, Huantong and Li, Xingwen},
title = {Time-resolved photon energy spectrum measurement of pulsed x rays from low-impedance rod-pinch diodes},
journal = {Review of Scientific Instruments},
year = {2026},
volume = {97},
number = {11},
doi = {10.1063/5.0351726}
}
EUVIMED