Impact of instrumental effects and target thickness on the SPARC magnetic proton recoil spectrometer response function

M Dalla Rosa iD, D Rigamonti iD, S Mackie iD, M Nocente iD, B I Buschmann iD, J Carmichael iD, J Kunimune iD, S Fugazza iD, S Colombi iD, A Dal Molin iD, E Gallo iD, G Gorini iD und 23 weitere

Review of Scientific Instruments · 01.11.2026 · 97 (11)

DOI: 10.1063/5.0350623 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Review of Scientific Instruments, online erschienen am 04.11.2026. Verfasst von M Dalla Rosa und 34 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physics, University of Milano-Bicocca, Milan 20126, Italy. Die Arbeit nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The SPARC tokamak aims to demonstrate net fusion power gain in tokamaks (Q-fus > 1). To measure neutron emission, a neutron spectrometer based on the magnetic proton recoil technique is being developed. This diagnostic operates by converting collimated neutrons into recoil protons through elastic scattering on a polyethylene target. The recoiled protons are then separated based on their momentum by a magnetic field and subsequently detected by a hodoscope composed of plastic scintillation detectors coupled to photomultiplier tubes. This allows neutron energies in the 2–20 MeV range to be discriminated, though only neutrons in a narrower energy window (±25% around the selected central energy) can be measured. The instrument will contribute to the determination of the total fusion power and provide information on plasma parameters such as ion temperature and the presence of non-thermal fuel ions. In this study, the design rationale of the hodoscope is presented and the main instrumental effects that govern proton collection are characterized. The hodoscope is divided into three regions with different channel widths, depths, and lengths, jointly optimized to satisfy three coupled physical requirements: sufficient energy resolution at the 14.1 MeV DT peak, full proton containment up to 18.75 MeV, and limitation of the background through minimization of the detector volume. Three correction factors are evaluated through dedicated simulations: the vertical signal loss due to protons falling outside the scintillator rod length, the cross-talk between adjacent channels, and the background contribution per channel. A correction scheme combining all three factors is presented.

Autor:innen und Institutionen

  • Department of Physics, University of Milano-Bicocca, Milan 20126, Italy.
  • Institute for Plasma Science and Technology, National Research Council, Milan 20125, Italy.
  • Massachusetts Institute of Technology, Cambridge, Massachusetts 02139, USA.
  • Commonwealth Fusion Systems, Devens, Massachusetts 01434, USA.
  • Eni S.p.A, Piazzale Mattei, 1, 00144 Roma, Italy.
  • Department of Physics and Astronomy, University of Padova, Padova 35131, Italy.

ORCID-Profile: M Dalla Rosa, D Rigamonti, S Mackie, M Nocente, B I Buschmann, J Carmichael, J Kunimune, S Fugazza, S Colombi, A Dal Molin, E Gallo, G Gorini, D Kelley, M H Kushoro, M Gatu Johnson, G Marcer, E Panontin, E Perelli Cippo, P Raj, M Rebai, M Reinke, A Seltzman, F Scioscioli, S Shubin, L Silvestrin, L G Tedoldi, R A Tinguely, M Tunesi, S Vanellone, K B Woller, M Tardocchi

Zitieren

APA

Dalla Rosa, M., Rigamonti, D., Mackie, S., Nocente, M., Buschmann, B. I., Carmichael, J., Kunimune, J., Fugazza, S., Colombi, S., Dal Molin, A., Gallo, E., Gorini, G., Kelley, D., Kushoro, M. H., Johnson, M. G., Litchfield, M., Marcer, G., Panontin, E., Parisi, M., …, & Tardocchi, M. (2026). Impact of instrumental effects and target thickness on the SPARC magnetic proton recoil spectrometer response function. Review of Scientific Instruments, 97(11). https://doi.org/10.1063/5.0350623

Vancouver

Dalla Rosa M, Rigamonti D, Mackie S, Nocente M, Buschmann BI, Carmichael J, et al. Impact of instrumental effects and target thickness on the SPARC magnetic proton recoil spectrometer response function. Review of Scientific Instruments. 2026;97(11). doi:10.1063/5.0350623

BibTeX
@article{DallaRosa2026,
  author = {Dalla Rosa, M and Rigamonti, D and Mackie, S and Nocente, M and Buschmann, B I and Carmichael, J and Kunimune, J and Fugazza, S and Colombi, S and Dal Molin, A and Gallo, E and Gorini, G and Kelley, D and Kushoro, M H and Johnson, M Gatu and Litchfield, M and Marcer, G and Panontin, E and Parisi, M and Cippo, E Perelli and Raj, P and Rebai, M and Reinke, M and Seltzman, A and Scioscioli, F and Shubin, S and Silvestrin, L and Tessaro, M and Tedoldi, L G and Tinguely, R A and Tunesi, M and Vanellone, S and Woller, K B and Woltjer, R and Tardocchi, M},
  title = {Impact of instrumental effects and target thickness on the SPARC magnetic proton recoil spectrometer response function},
  journal = {Review of Scientific Instruments},
  year = {2026},
  volume = {97},
  number = {11},
  doi = {10.1063/5.0350623}
}
Kostenlos starten →