Digital least-squares amplitude detection for high-speed atomic force microscopy
Review of Scientific Instruments · 01.11.2026 · 97 (11)
DOI: 10.1063/5.0348515 ↗Kurz gefasst
Fachartikel in der Zeitschrift Review of Scientific Instruments, online erschienen am 16.11.2026. Verfasst von Suresh Babu Lavanya und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Chemistry, University of Texas at Austin, Austin, Texas 78712, USA. Die Arbeit nennt 20 Literaturangaben. Gefördert von: University of Texas at Austin, National Institute of General Medical Sciences.
Abstract (Originalsprache)
High-speed atomic force microscopy (HS-AFM) in amplitude-modulated (tapping) mode requires rapid and reliable amplitude detection for real-time feedback control. Here, we develop a low-latency field-programmable gate array (FPGA)-based digital least-squares (LSQ) amplitude detector for HS-AFM operation. Although LSQ is a well-established signal-processing method, the novelty of this work lies in its implementation as a high-frequency, microsecond-response amplitude detector for real-time HS-AFM feedback. To enable real-time operation, we implemented a two-parameter LSQ formulation on FPGA hardware, in which the adjustable fitting-window length defines the number of sampled points used for each amplitude estimate. We evaluated LSQ using numerical simulations, signal-generator measurements, HS-AFM deflection-signal analysis, and biological HS-AFM imaging. Under matched response-time conditions, numerical simulations showed that LSQ produced amplitude-estimate fluctuations comparable to Fourier-amplitude-based detection and lower than those produced by sample-and-hold-based detection (SHB) and phase-shift-based detection. Experimental measurements further showed that shorter LSQ windows provided SHB-like response times, whereas longer windows reduced amplitude-estimate fluctuations. LSQ also supported stable HS-AFM imaging of annexin-V lattices on supported lipid bilayers. These results establish digital LSQ as an additional amplitude-detection strategy for HS-AFM, particularly for applications involving noisy cantilever deflection signals where tunable control of the response-time/amplitude-fluctuation trade-off is needed.
Autor:innen und Institutionen
- Department of Chemistry, University of Texas at Austin, Austin, Texas 78712, USA.
ORCID-Profile: Suresh Babu Lavanya, Yi-Chih Lin
Förderung
- University of Texas at Austin
- National Institute of General Medical Sciences
Zitieren
APA
Lavanya, S. B., & Lin, Y. C. (2026). Digital least-squares amplitude detection for high-speed atomic force microscopy. Review of Scientific Instruments, 97(11). https://doi.org/10.1063/5.0348515
Vancouver
Lavanya SB, Lin YC. Digital least-squares amplitude detection for high-speed atomic force microscopy. Review of Scientific Instruments. 2026;97(11). doi:10.1063/5.0348515
BibTeX
@article{Lavanya2026,
author = {Lavanya, Suresh Babu and Lin, Yi-Chih},
title = {Digital least-squares amplitude detection for high-speed atomic force microscopy},
journal = {Review of Scientific Instruments},
year = {2026},
volume = {97},
number = {11},
doi = {10.1063/5.0348515}
}
EUVIMED