EUVIMEDEUVIMED entdecken
Crossref

Development of a cavity-enhanced absorption spectroscopy based heavy water leak monitor for nuclear reactors

Anita Gupta iD, Sanjay P Somkuwar iD, Yogesh G Arondekar, Aparna Shastri iD, R Sampathkumar, Dinesh V Udupa iD

Review of Scientific Instruments · 01.11.2026 · 97 (11)

DOI: 10.1063/5.0336753 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Review of Scientific Instruments, online erschienen am 02.11.2026. Verfasst von Anita Gupta und 5 weiteren Autor:innen. Beteiligte Einrichtung: Atomic and Molecular Physics Division, Bhabha Atomic Research Centre, Mumbai, India. Die Arbeit nennt 47 Literaturangaben.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The development of a compact, low-cost instrument for reliable, continuous detection of leaks of heavy water in pressurized heavy water reactors (PHWRs) by selectively detecting semi-heavy water (HDO) in air is presented. The device employs cavity-enhanced absorption spectroscopy by targeting a specific rovibrational transition of HDO to achieve high sensitivity and selectivity. The complete development of the cavity-enhanced absorption spectroscopy based HDO in air monitor (CHArM), from laboratory feasibility studies to a field-deployable prototype, is described. This includes the optical design, instrument configuration, calibration methodology, and performance evaluation. Major technical challenges faced during development and the solutions implemented for stable, long-term operation are discussed. The CHArM instrument has achieved a measurement precision of ±10.6 ppbv for HDO in air, which is a considerable improvement over the best sensitivities reported for HDO detection instruments used in PHWRs. It offers a simpler, cost-effective, and easily deployable solution for continuous heavy water leak monitoring in PHWR settings.

Autor:innen und Institutionen

  • Atomic and Molecular Physics Division, Bhabha Atomic Research Centre, Mumbai, India.
  • Homi Bhabha National Institute, Mumbai, India.
  • Information Management and Coordination Division, Department of Atomic Energy, Mumbai, India.

ORCID-Profile: Anita Gupta, Sanjay P Somkuwar, Aparna Shastri, Dinesh V Udupa

Zitieren

APA

Gupta, A., Somkuwar, S. P., Arondekar, Y. G., Shastri, A., Sampathkumar, R., & Udupa, D. V. (2026). Development of a cavity-enhanced absorption spectroscopy based heavy water leak monitor for nuclear reactors. Review of Scientific Instruments, 97(11). https://doi.org/10.1063/5.0336753

Vancouver

Gupta A, Somkuwar SP, Arondekar YG, Shastri A, Sampathkumar R, Udupa DV. Development of a cavity-enhanced absorption spectroscopy based heavy water leak monitor for nuclear reactors. Review of Scientific Instruments. 2026;97(11). doi:10.1063/5.0336753

BibTeX
@article{Gupta2026,
  author = {Gupta, Anita and Somkuwar, Sanjay P and Arondekar, Yogesh G and Shastri, Aparna and Sampathkumar, R and Udupa, Dinesh V},
  title = {Development of a cavity-enhanced absorption spectroscopy based heavy water leak monitor for nuclear reactors},
  journal = {Review of Scientific Instruments},
  year = {2026},
  volume = {97},
  number = {11},
  doi = {10.1063/5.0336753}
}
Kostenlos starten →