EUVIMEDEUVIMED entdecken
Crossref

Design and implementation of a thermoelectric cooler-based temperature control system for the resonator tube wall of an optically pumped terahertz gas laser

Zhiyong Zou iD, Jianhang Zhao, Jiaxing Xie iD, Shuai Wu, Haishun Deng, Nu Zhang, Anlan Huang, Xiang Ma, Huanhuan Tian

Review of Scientific Instruments · 01.05.2026 · 97 (5)

DOI: 10.1063/5.0320955 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Review of Scientific Instruments, online erschienen am 29.05.2026. Verfasst von Zhiyong Zou und 8 weiteren Autor:innen. Beteiligte Einrichtung: School of Mechanical and Electrical Engineering, Anhui University of Science and Technology, Huainan 232001, China. Die Arbeit nennt 16 Literaturangaben. Gefördert von: National Natural Science Foundation of China, Institute of Energy, Hefei Comprehensive National Science Center.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

A thermoelectric cooler (TEC)-based temperature control system is developed for the resonator tube wall of an optically pumped terahertz gas laser, which serves as the light source of a polarimeter–interferometer system applied in tokamak device diagnostics. The core objective of this system is to ensure the continuous and stable measurement performance of the polarimeter–interferometer during plasma diagnosis. The thermal module is composed of aluminum finned heat sinks that fully enclose the resonator tube wall, within which is embedded a semiconductor cooling element (TEC) and a PT1000 platinum resistance thermometer. The implementation of a fuzzy proportional–integral–derivative (PID) algorithm by a Field-Programmable Gate Array (FPGA)-based controller serves to regulate both the magnitude and polarity of the TEC drive current. This, in turn, maintains the temperature of the cavity wall. In the experiment, the temperature of the tube wall was maintained at 15 ± 0.25 °C. Concurrently, the laser output power was stabilized at 13 ± 0.2 mW. In addition, the unloaded thermal module displayed a temperature fluctuation of ±0.1 °C. The compact, high-precision design facilitates laser miniaturization and underlies the accuracy of interferometric plasma density diagnostics.

Autor:innen und Institutionen

  • School of Mechanical and Electrical Engineering, Anhui University of Science and Technology, Huainan 232001, China.
  • Institute of Energy, Hefei Comprehensive National Science Center, Hefei 230031, Anhui, China.
  • School of Electronic and Information Engineering, Anhui University, Hefei 230039, China.
  • Institute of Plasma Physics, Hefei Institutes of Physical Science, Chinese Academy of Sciences, Hefei 230031, China.

ORCID-Profile: Zhiyong Zou, Jiaxing Xie

Förderung

  • National Natural Science Foundation of China
  • Institute of Energy, Hefei Comprehensive National Science Center

Zitieren

APA

Zou, Z., Zhao, J., Xie, J., Wu, S., Deng, H., Zhang, N., Huang, A., Ma, X., & Tian, H. (2026). Design and implementation of a thermoelectric cooler-based temperature control system for the resonator tube wall of an optically pumped terahertz gas laser. Review of Scientific Instruments, 97(5). https://doi.org/10.1063/5.0320955

Vancouver

Zou Z, Zhao J, Xie J, Wu S, Deng H, Zhang N, et al. Design and implementation of a thermoelectric cooler-based temperature control system for the resonator tube wall of an optically pumped terahertz gas laser. Review of Scientific Instruments. 2026;97(5). doi:10.1063/5.0320955

BibTeX
@article{Zou2026,
  author = {Zou, Zhiyong and Zhao, Jianhang and Xie, Jiaxing and Wu, Shuai and Deng, Haishun and Zhang, Nu and Huang, Anlan and Ma, Xiang and Tian, Huanhuan},
  title = {Design and implementation of a thermoelectric cooler-based temperature control system for the resonator tube wall of an optically pumped terahertz gas laser},
  journal = {Review of Scientific Instruments},
  year = {2026},
  volume = {97},
  number = {5},
  doi = {10.1063/5.0320955}
}