Lateral force microscopy calibration using an interferometric atomic force microscope
Review of Scientific Instruments · 01.04.2026 · 97 (4)
DOI: 10.1063/5.0318278 ↗Kurz gefasst
Fachartikel in der Zeitschrift Review of Scientific Instruments, online erschienen am 21.04.2026. Verfasst von Joel A Lefever und 2 weiteren Autor:innen. Beteiligte Einrichtung: Oxford Instruments Asylum Research, Inc., Santa Barbara, CA, 93117. Die Arbeit nennt 25 Literaturangaben.
Abstract (Originalsprache)
A new method is introduced for calibrating lateral force as measured by an atomic force microscope, making use of both an interferometric detector and an optical beam detector on the same instrument. The method may be implemented automatically and performed with minimal user input. The microscope has the capability to measure the probe tip height in situ, which allows for a complete lateral force calibration without changing the sample or probe. Two options for measuring lateral force are described, wherein the two detectors are alternately used to measure normal and lateral forces, and methods for applying the calibration protocol for both alternatives are provided. The tip height measurement is validated by direct comparison with an electron micrograph, and the two methods are generally in agreement to within 1.4 μm. For most cantilevers tested, the complete lateral calibration method is consistent with the wedge calibration method within the intrinsic uncertainty of the wedge method.
Autor:innen und Institutionen
- Oxford Instruments Asylum Research, Inc., Santa Barbara, CA, 93117.
ORCID-Profile: Joel A Lefever, Aleksander Labuda, Roger Proksch
Zitieren
APA
Lefever, J. A., Labuda, A., & Proksch, R. (2026). Lateral force microscopy calibration using an interferometric atomic force microscope. Review of Scientific Instruments, 97(4). https://doi.org/10.1063/5.0318278
Vancouver
Lefever JA, Labuda A, Proksch R. Lateral force microscopy calibration using an interferometric atomic force microscope. Review of Scientific Instruments. 2026;97(4). doi:10.1063/5.0318278
BibTeX
@article{Lefever2026,
author = {Lefever, Joel A and Labuda, Aleksander and Proksch, Roger},
title = {Lateral force microscopy calibration using an interferometric atomic force microscope},
journal = {Review of Scientific Instruments},
year = {2026},
volume = {97},
number = {4},
doi = {10.1063/5.0318278}
}