Micro-computed tomography beamline of the Australian synchrotron: Density measurements

Benedicta D Arhatari iD, Darren Thompson iD, Andrew W Stevenson iD, Adam Walsh, Tom Fiala iD, Gary Ruben iD, Tingting Feng, Stephen Mudie

Review of Scientific Instruments · 01.02.2025 · 96 (2)

DOI: 10.1063/5.0245931 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Review of Scientific Instruments, online erschienen am 26.02.2025. Verfasst von Benedicta D Arhatari und 7 weiteren Autor:innen. Beteiligte Einrichtung: Australian Synchrotron, ANSTO, Clayton, Victoria 3168, Australia. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 40 Literaturangaben.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The Micro-Computed Tomography (MCT) beamline at the Australian Synchrotron (ANSTO) offers superior capabilities in micrometer-scale spatial resolution and three-dimensional x-ray imaging. MCT is the first of the eight new BRIGHT beamlines and has been operating successfully with users for approximately two years. It is a bending magnet beamline capable of delivering a white beam, a pink beam, or a monochromatic beam in the 8–40 keV energy range using a Double Multilayer Monochromator (DMM). Ongoing development continues at the MCT beamline to extend its capabilities. In this article, we present the operation and energy calibration of the DMM, highlighting the unique advantages offered by synchrotron-based micro-CT and its application for quantitative imaging, such as density measurements using monochromatic energy.

Autor:innen und Institutionen

  • Australian Synchrotron, ANSTO, Clayton, Victoria 3168, Australia.
  • Department of Mathematical and Physical Sciences, La Trobe University, Bundoora, Victoria 3086, Australia.
  • CSIRO, Clayton, Victoria 3168, Australia.

ORCID-Profile: Benedicta D Arhatari, Darren Thompson, Andrew W Stevenson, Tom Fiala, Gary Ruben

Zitieren

APA

Arhatari, B. D., Thompson, D., Stevenson, A. W., Walsh, A., Fiala, T., Ruben, G., Feng, T., & Mudie, S. (2025). Micro-computed tomography beamline of the Australian synchrotron: Density measurements. Review of Scientific Instruments, 96(2). https://doi.org/10.1063/5.0245931

Vancouver

Arhatari BD, Thompson D, Stevenson AW, Walsh A, Fiala T, Ruben G, et al. Micro-computed tomography beamline of the Australian synchrotron: Density measurements. Review of Scientific Instruments. 2025;96(2). doi:10.1063/5.0245931

BibTeX
@article{Arhatari2025,
  author = {Arhatari, Benedicta D and Thompson, Darren and Stevenson, Andrew W and Walsh, Adam and Fiala, Tom and Ruben, Gary and Feng, Tingting and Mudie, Stephen},
  title = {Micro-computed tomography beamline of the Australian synchrotron: Density measurements},
  journal = {Review of Scientific Instruments},
  year = {2025},
  volume = {96},
  number = {2},
  doi = {10.1063/5.0245931}
}
Kostenlos starten →