Measuring the thermal conductivity of hydrogels with a bidirectional 3ω method
Review of Scientific Instruments · 01.04.2025 · 96 (4)
DOI: 10.1063/5.0245268 ↗Kurz gefasst
Fachartikel in der Zeitschrift Review of Scientific Instruments, online erschienen am 14.04.2025. Verfasst von Daniel H Hsieh und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Mechanical Science and Engineering, University of Illinois Urbana-Champaign, 1206 W Green St., Urbana, Illinois 61801, USA. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 56 Literaturangaben. Gefördert von: U.S. Department of Energy EERE Building Technologies Office.
Abstract (Originalsprache)
Hydrogels are soft, water-absorbing polymer materials with diverse applications in biomedicine and agriculture. Recently, hydrogels have been proposed to encapsulate water-soluble phase change materials that store energy in their latent heat of solidification. In these applications, the thermal conductivity of these materials affects their performance. Few methods exist for measuring the thermal conductivity of small quantities of hydrogels. Here, we describe an implementation of the bidirectional 3ω technique to measure the thermal conductivity of hydrogels with particular attention to their moisture content. Our implementation of the technique can probe sample volumes as little as ∼20 μl and yields the thermal conductivity without requiring fitting of additional thermal parameters. We numerically simulate 3ω sensor designs with frequency-domain 3-D models to quantify and reduce errors introduced by the choice of substrate and insulation layer thickness. Frequencies in the ∼1–20 Hz range yield less error for the materials considered here. We verify our setup with measurements on water and report values for polyacrylamide and poly(2-acrylamido-2-methylpropane sulfonic acid) (PAMPS) hydrogels. Our swollen hydrogels exhibited thermal conductivities nearly equivalent to water, 0.6 W m−1 K−1, and we estimate thermal conductivities of 0.43 and 0.42 W m−1 K−1 for neat polyacrylamide and PAMPS, respectively. Finally, we estimate an error of ±7%, consistent with other 3ω methods, with the largest error coming from the sensor calibration. We find that our implementation of the bidirectional 3ω method gives reasonable results and can be employed for prototyping soft materials relevant to thermal storage.
Autor:innen und Institutionen
- Department of Mechanical Science and Engineering, University of Illinois Urbana-Champaign, 1206 W Green St., Urbana, Illinois 61801, USA.
- Department of Materials Science and Engineering, University of Illinois Urbana-Champaign, 1304 W Green St., Urbana, Illinois 61801, USA.
- Materials Research Laboratory, University of Illinois Urbana-Champaign, 104 S Goodwin Ave., Urbana, Illinois 61801, USA.
- Holonyak Micro and Nanotechnology Laboratory, University of Illinois Urbana-Champaign, 208 N Wright St., Urbana, Illinois 61801, USA.
ORCID-Profile: Daniel H Hsieh, Youngmun Lee, Mayur S Prabhudesai, Jay Taylor, Paul V Braun, Sanjiv Sinha
Förderung
- U.S. Department of Energy EERE Building Technologies Office
Zitieren
APA
Hsieh, D. H., Lee, Y., Prabhudesai, M. S., Taylor, J., Braun, P. V., & Sinha, S. (2025). Measuring the thermal conductivity of hydrogels with a bidirectional 3ω method. Review of Scientific Instruments, 96(4). https://doi.org/10.1063/5.0245268
Vancouver
Hsieh DH, Lee Y, Prabhudesai MS, Taylor J, Braun PV, Sinha S. Measuring the thermal conductivity of hydrogels with a bidirectional 3ω method. Review of Scientific Instruments. 2025;96(4). doi:10.1063/5.0245268
BibTeX
@article{Hsieh2025,
author = {Hsieh, Daniel H and Lee, Youngmun and Prabhudesai, Mayur S and Taylor, Jay and Braun, Paul V and Sinha, Sanjiv},
title = {Measuring the thermal conductivity of hydrogels with a bidirectional 3ω method},
journal = {Review of Scientific Instruments},
year = {2025},
volume = {96},
number = {4},
doi = {10.1063/5.0245268}
}
EUVIMED