Fachartikel in der Zeitschrift APL Photonics, online erschienen am 23.10.2024. Verfasst von Jake R Rosvold und 4 weiteren Autor:innen. Beteiligte Einrichtung: Fischell Department of Bioengineering, University of Maryland, 8278 Paint Branch Drive, College Park, Maryland 20742, USA. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: NCI NIH HHS, U.S. Naval Research Laboratory, National Institutes of Health.
Abstract (Originalsprache)
Brillouin spectroscopy has become an important tool for mapping the mechanical properties of biological samples. Recently, stimulated Brillouin scattering (SBS) measurements have emerged in this field as a promising technology for lower noise and higher speed measurements. However, further improvements are fundamentally limited by constraints on the optical power level that can be used in biological samples, which effectively caps the gain and signal-to-noise ratio (SNR) of SBS biological measurements. This limitation is compounded by practical limits on the optical probe power due to detector saturation thresholds. As a result, SBS-based measurements in biological samples have provided minimal improvements (in noise and imaging speed) compared with spontaneous Brillouin microscopy, despite the potential advantages of the nonlinear scattering process. Here, we consider how a SBS spectrometer can circumvent this fundamental trade-off in the low-gain regime by leveraging the polarization dependence of the SBS interaction to effectively filter the signal from the background light via the polarization pulling effect. We present an analytic model of the polarization pulling detection scheme and describe the trade-space unique to Brillouin microscopy applications. We show that an optimized receiver design could provide >25× improvement in SNR compared to a standard SBS receiver in most typical experimental conditions. We then experimentally validate this model using optical fiber as a simplified test bed. With our experimental parameters, we find that the polarization pulling scheme provides 100× higher SNR than a standard SBS receiver, enabling 100× faster measurements in the low-gain regime. Finally, we discuss the potential for this proposed spectrometer design to benefit low-gain spectroscopy applications such as Brillouin microscopy by enabling pixel dwell times as short as 10 μs.
Autor:innen und Institutionen
- Fischell Department of Bioengineering, University of Maryland, 8278 Paint Branch Drive, College Park, Maryland 20742, USA.
- U.S. Naval Research Laboratory, 4555 Overlook Ave. SW, Washington, District of Columbia 20375, USA.
ORCID-Profile: Jake R Rosvold, Joseph B Murray, Giulia Zanini, Brandon Redding, Giuliano Scarcelli
Förderung
- NCI NIH HHS
- U.S. Naval Research Laboratory
- National Institutes of Health
Zitieren
APA
Rosvold, J. R., Murray, J. B., Zanini, G., Redding, B., & Scarcelli, G. (2024). Impact of polarization pulling on optimal spectrometer design for stimulated Brillouin scattering microscopy. APL Photonics, 9(10). https://doi.org/10.1063/5.0225074
Vancouver
Rosvold JR, Murray JB, Zanini G, Redding B, Scarcelli G. Impact of polarization pulling on optimal spectrometer design for stimulated Brillouin scattering microscopy. APL Photonics. 2024;9(10). doi:10.1063/5.0225074
BibTeX
@article{Rosvold2024,
author = {Rosvold, Jake R and Murray, Joseph B and Zanini, Giulia and Redding, Brandon and Scarcelli, Giuliano},
title = {Impact of polarization pulling on optimal spectrometer design for stimulated Brillouin scattering microscopy},
journal = {APL Photonics},
year = {2024},
volume = {9},
number = {10},
doi = {10.1063/5.0225074}
}
EUVIMED