Design of a Thomson scattering diagnostic for the SMall Aspect Ratio Tokamak (SMART)

M Kaur iD, A Diallo iD, B LeBlanc iD, J Segado-Fernandez, E Viezzer iD, R B Huxford, A Mancini iD, D J Cruz-Zabala iD, M Podesta iD, J W Berkery iD, M Garcia-Muñoz iD

Review of Scientific Instruments · 01.09.2024 · 95 (9)

DOI: 10.1063/5.0219308 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Review of Scientific Instruments, online erschienen am 04.09.2024. Verfasst von M Kaur und 10 weiteren Autor:innen. Beteiligte Einrichtung: Princeton Plasma Physics Laboratory, Princeton, New Jersey 08540, USA. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 20 Literaturangaben. Gefördert von: U.S. Department of Energy, EUROfusion.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

We describe the design of a Thomson scattering (TS) diagnostic to be used on the SMall Aspect Ratio Tokamak (SMART). SMART is a spherical tokamak being commissioned in Spain that aims to explore positive triangularity and negative triangularity plasma scenarios at a low aspect ratio. The SMART TS diagnostic is designed to operate at high spatial resolution, 6 mm scattering length in the low-field side and 9 mm in the high-field side regions, and a wide dynamic range, electron temperature from 1 eV to 1 keV and density from 5×1018m−3 to 1×1020m−3, to resolve large gradients formed at the plasma edge and in the scrape-off layer (SOL) under different triangularities and low aspect ratios. A 2 J @1064 nm laser will be used that is capable of operating in the burst mode at 1, 2, and 4 kHz to investigate fast phenomena and at 30 Hz to study 1 s (or more) long discharges. The scattered light will be collected over an angular range of 60° – 120° from 28 spatial points in the midplane covering the entire plasma width and the outer midplane SOL. Each scattering signal will be spectrally resolved on five wavelength channels of a polychromator to obtain the electron temperature measurement. We will also present a method to monitor in situ laser alignment in the core during calibrations and plasma operations.

Autor:innen und Institutionen

  • Princeton Plasma Physics Laboratory, Princeton, New Jersey 08540, USA.
  • Department of Atomic, Molecular and Nuclear Physics, University of Seville, Seville, Spain.
  • RBH Optics, Burgess Hill, West Sussex RH15 8HL, United Kingdom.

ORCID-Profile: M Kaur, A Diallo, B LeBlanc, E Viezzer, A Mancini, D J Cruz-Zabala, M Podesta, J W Berkery, M Garcia-Muñoz

Förderung

  • U.S. Department of Energy
  • EUROfusion

Zitieren

APA

Kaur, M., Diallo, A., LeBlanc, B., Segado-Fernandez, J., Viezzer, E., Huxford, R. B., Mancini, A., Cruz-Zabala, D. J., Podesta, M., Berkery, J. W., & Garcia-Muñoz, M. (2024). Design of a Thomson scattering diagnostic for the SMall Aspect Ratio Tokamak (SMART). Review of Scientific Instruments, 95(9). https://doi.org/10.1063/5.0219308

Vancouver

Kaur M, Diallo A, LeBlanc B, Segado-Fernandez J, Viezzer E, Huxford RB, et al. Design of a Thomson scattering diagnostic for the SMall Aspect Ratio Tokamak (SMART). Review of Scientific Instruments. 2024;95(9). doi:10.1063/5.0219308

BibTeX
@article{Kaur2024,
  author = {Kaur, M and Diallo, A and LeBlanc, B and Segado-Fernandez, J and Viezzer, E and Huxford, R B and Mancini, A and Cruz-Zabala, D J and Podesta, M and Berkery, J W and Garcia-Muñoz, M},
  title = {Design of a Thomson scattering diagnostic for the SMall Aspect Ratio Tokamak (SMART)},
  journal = {Review of Scientific Instruments},
  year = {2024},
  volume = {95},
  number = {9},
  doi = {10.1063/5.0219308}
}
Kostenlos starten →