Electrical design of the flexible imaging diffraction diagnostic for laser experiments (FIDDLE) at the National Ignition Facility (NIF)—Requirements, design, and performance

P R Nyholm iD, N E Palmer iD, R B Petre, A C Carpenter, S R Nagel iD, D K Bradley iD, C E Vennari iD, B Golick iD, D Boyle, A M Sharp iD, S B Morioka iD, A J Mackinnon iD und 1 weitere

Review of Scientific Instruments · 01.07.2024 · 95 (7)

DOI: 10.1063/5.0216863 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Review of Scientific Instruments, online erschienen am 23.07.2024. Verfasst von P R Nyholm und 12 weiteren Autor:innen. Beteiligte Einrichtung: Lawrence Livermore National Laboratory, Livermore, California 94550, USA. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 9 Literaturangaben. Gefördert von: U.S. Department of Energy.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The Flexible Imaging Diffraction Diagnostic for Laser Experiments (FIDDLE) is a newly developed diagnostic for imaging time resolved diffraction in experiments at the National Ignition Facility (NIF). It builds on the successes of its predecessor, the Gated Diffraction Development Diagnostic (G3D). The FIDDLE was designed to support eight Daedalus version 2 sensors (six more hCMOS sensors than any other hCMOS-based diagnostic in NIF to date) and an integrated streak camera. We will review the electrical requirements, design, and performance of the electrical subsystems that were created to support this large number of cameras in the FIDDLE. The analysis of the data that the FIDDLE is intended to collect relies heavily on the accurate and well-understood timing of each sensor. We report camera-to-camera timing jitter of less than 100 ps rms and sensor integration times of 2.2 ns FWHM in 2-2 timing mode. Additionally, diffraction experiments on the NIF produce electric fields (EMI) on the order of 1 kV/m, which have been observed to negatively impact the performance of some electrical components of the FIDDLE. We report on the results of testing hCMOS camera electronics in a similar EMI environment generated in an offline lab. We also summarize the use of a novel approach to using a vector network analyzer as an EMI leak detector to understand and reduce the negative impacts of EMI on the FIDDLE.

Autor:innen und Institutionen

  • Lawrence Livermore National Laboratory, Livermore, California 94550, USA.

ORCID-Profile: P R Nyholm, N E Palmer, S R Nagel, D K Bradley, C E Vennari, B Golick, A M Sharp, S B Morioka, A J Mackinnon, L R Benedetti

Förderung

  • U.S. Department of Energy

Zitieren

APA

Nyholm, P. R., Palmer, N. E., Petre, R. B., Carpenter, A. C., Nagel, S. R., Bradley, D. K., Vennari, C. E., Golick, B., Boyle, D., Sharp, A. M., Morioka, S. B., Mackinnon, A. J., & Benedetti, L. R. (2024). Electrical design of the flexible imaging diffraction diagnostic for laser experiments (FIDDLE) at the National Ignition Facility (NIF)—Requirements, design, and performance. Review of Scientific Instruments, 95(7). https://doi.org/10.1063/5.0216863

Vancouver

Nyholm PR, Palmer NE, Petre RB, Carpenter AC, Nagel SR, Bradley DK, et al. Electrical design of the flexible imaging diffraction diagnostic for laser experiments (FIDDLE) at the National Ignition Facility (NIF)—Requirements, design, and performance. Review of Scientific Instruments. 2024;95(7). doi:10.1063/5.0216863

BibTeX
@article{Nyholm2024,
  author = {Nyholm, P R and Palmer, N E and Petre, R B and Carpenter, A C and Nagel, S R and Bradley, D K and Vennari, C E and Golick, B and Boyle, D and Sharp, A M and Morioka, S B and Mackinnon, A J and Benedetti, L R},
  title = {Electrical design of the flexible imaging diffraction diagnostic for laser experiments (FIDDLE) at the National Ignition Facility (NIF)—Requirements, design, and performance},
  journal = {Review of Scientific Instruments},
  year = {2024},
  volume = {95},
  number = {7},
  doi = {10.1063/5.0216863}
}
Kostenlos starten →