Monte Carlo toolkit for designing and validating step-range-filter spectrometer designs
Review of Scientific Instruments · 01.02.2025 · 96 (2)
DOI: 10.1063/5.0206206 ↗Kurz gefasst
Fachartikel in der Zeitschrift Review of Scientific Instruments, online erschienen am 04.02.2025. Verfasst von T M Johnson und 19 weiteren Autor:innen. Beteiligte Einrichtung: Massachusetts Institute of Technology, Cambridge, Massachusetts 02139, USA. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Nuclear Security Administration.
Abstract (Originalsprache)
Here, we present a Monte Carlo toolkit for validating step range filter (SRF) spectrometer designs. Geant4 is used to transport charged particles through the SRF filters to generate synthetic SRF data that include realistic CR-39 effects. Synthetic SRF spectra generated by this method inherently account for instrument response and allow for the quantification of SRF performance before shots. The usefulness of this toolkit is demonstrated through its application to a number of problems. A new broadband SRF for the ∼10 MeV wide 3He3He proton spectrum is validated, and an analysis method for analyzing 3He3He-p SRF data that accounts for instrument response is put forth. In addition, an SRF design for the compact recoil-proton spectrometer (CRS) on the Z-machine is validated. Finally, a new calibration technique for the DD-p SRF is proposed and validated.
Autor:innen und Institutionen
- Massachusetts Institute of Technology, Cambridge, Massachusetts 02139, USA.
- Lawrence Livermore National Laboratory, Livermore, California 94550, USA.
- Laboratory for Laser Energetics, University of Rochester, Rochester, New York 14623, USA.
- Los Alamos National Laboratory, Los Alamos, New Mexico 87545, USA.
ORCID-Profile: T M Johnson, B Lahmann, L Russell, N L Vanderloo, M J Cufari, B L Reichelt, C W Chang, A Birkel, N V Kabadi, G D Sutcliffe, P J Adrian, J A Pearcy, J H Kunimune, S G Dannhoff, T E Evans, M Gatu Johnson, F H Séguin, R D Petrasso, C K Li, J A Frenje
Förderung
- National Nuclear Security Administration
Zitieren
APA
Johnson, T. M., Lahmann, B., Russell, L., Vanderloo, N. L., Cufari, M. J., Reichelt, B. L., Chang, C. W., Birkel, A., Kabadi, N. V., Sutcliffe, G. D., Adrian, P. J., Pearcy, J. A., Kunimune, J. H., Dannhoff, S. G., Evans, T. E., Johnson, M. G., Séguin, F. H., Petrasso, R. D., Li, C. K., & Frenje, J. A. (2025). Monte Carlo toolkit for designing and validating step-range-filter spectrometer designs. Review of Scientific Instruments, 96(2). https://doi.org/10.1063/5.0206206
Vancouver
Johnson TM, Lahmann B, Russell L, Vanderloo NL, Cufari MJ, Reichelt BL, et al. Monte Carlo toolkit for designing and validating step-range-filter spectrometer designs. Review of Scientific Instruments. 2025;96(2). doi:10.1063/5.0206206
BibTeX
@article{Johnson2025,
author = {Johnson, T M and Lahmann, B and Russell, L and Vanderloo, N L and Cufari, M J and Reichelt, B L and Chang, C W and Birkel, A and Kabadi, N V and Sutcliffe, G D and Adrian, P J and Pearcy, J A and Kunimune, J H and Dannhoff, S G and Evans, T E and Johnson, M Gatu and Séguin, F H and Petrasso, R D and Li, C K and Frenje, J A},
title = {Monte Carlo toolkit for designing and validating step-range-filter spectrometer designs},
journal = {Review of Scientific Instruments},
year = {2025},
volume = {96},
number = {2},
doi = {10.1063/5.0206206}
}
EUVIMED