Degaussing procedure and performance enhancement by low-frequency shaking of a 3-layer magnetically shielded room
Review of Scientific Instruments · 01.11.2023 · 94 (11)
DOI: 10.1063/5.0167663 ↗Kurz gefasst
Fachartikel in der Zeitschrift Review of Scientific Instruments, online erschienen am 07.11.2023. Verfasst von Fabian Allmendinger und 3 weiteren Autor:innen. Beteiligte Einrichtung: Physikalisches Institut, Universität Heidelberg, 69120 Heidelberg, Germany. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 31 Literaturangaben. Gefördert von: Deutsche Forschungsgemeinschaft.
Abstract (Originalsprache)
We report on the performance of a Magnetically Shielded Room (MSR) intended for next level 3He/129Xe co-magnetometer experiments that require improved magnetic conditions. The MSR consists of three layers of Mu-metal with a thickness of 3 mm each and one additional highly conductive copper-coated aluminum layer with a thickness of 8 mm. It has a cubical shape with a walk-in interior volume with an edge length of 2560 mm. An optimized degaussing (magnetic equilibration) procedure using a frequency sweep with a constant amplitude followed by an exponential decay of the amplitude will be presented. The procedure for the whole MSR takes 21 min, and measurements of the residual magnetic field at the center of the MSR show that |B| < 1 nT can be reached reliably. The chosen degaussing procedure will be motivated by online hysteresis measurements of the assembled MSR and by eddy-current simulations, showing that saturation at the center of the Mu-metal layer is reached. Shielding factors can be improved by a factor ≈4 in all directions by low frequency (0.2 Hz), low current (1 A) shaking of the outermost Mu-metal layer.
Autor:innen und Institutionen
- Physikalisches Institut, Universität Heidelberg, 69120 Heidelberg, Germany.
- Institut für Physik, Universität Mainz, 55128 Mainz, Germany.
ORCID-Profile: Fabian Allmendinger, Benjamin Brauneis, Werner Heil, Ulrich Schmidt
Förderung
- Deutsche Forschungsgemeinschaft
Zitieren
APA
Allmendinger, F., Brauneis, B., Heil, W., & Schmidt, U. (2023). Degaussing procedure and performance enhancement by low-frequency shaking of a 3-layer magnetically shielded room. Review of Scientific Instruments, 94(11). https://doi.org/10.1063/5.0167663
Vancouver
Allmendinger F, Brauneis B, Heil W, Schmidt U. Degaussing procedure and performance enhancement by low-frequency shaking of a 3-layer magnetically shielded room. Review of Scientific Instruments. 2023;94(11). doi:10.1063/5.0167663
BibTeX
@article{Allmendinger2023,
author = {Allmendinger, Fabian and Brauneis, Benjamin and Heil, Werner and Schmidt, Ulrich},
title = {Degaussing procedure and performance enhancement by low-frequency shaking of a 3-layer magnetically shielded room},
journal = {Review of Scientific Instruments},
year = {2023},
volume = {94},
number = {11},
doi = {10.1063/5.0167663}
}
EUVIMED