Microbiological Monitoring of Endoscopes: A Single-Center Observational Study
Endoscopy · 01.06.2026 · 58 (S 03 S 03) · S35
DOI: 10.1055/s-0046-1820693 ↗Kurz gefasst
Fachartikel in der Zeitschrift Endoscopy, erschienen am 01.06.2026. Verfasst von J Dolcini und 12 weiteren Autor:innen.
Abstract (Originalsprache)
Aims
Preventing endoscopy-related infections is a key patient-safety priority. Although reprocessing procedures are regulated by international guidelines [ 1 ], persistent contamination has been documented—particularly in duodenoscopes—due to their structural complexity [ 2 ]. Periodic microbiological monitoring is an essential tool for verifying reprocessing efficacy and identifying potential critical issues [ 3 ].To evaluate the effectiveness of endoscope reprocessing through microbiological monitoring and to identify criticalities associated with specific endoscope types or individual instruments.
Methods
A single-center observational study was conducted, including all microbiological samples collected from January to September 2024 on endoscopes in clinical use. For each sample, the following data were recorded: instrument type and ID, sampling date, isolation of High-Concern Organisms (HCO) or Low-Concern Organisms (LCO), bacterial load (CFU), and corrective actions undertaken. Absolute and relative frequencies of positive samples were calculated, stratified by endoscope type and by individual instrument. Descriptive statistical analyses were performed using R (version 4.2.3; R Foundation for Statistical Computing, Vienna, Austria).
Results
A total of 71 samples were collected from 39 endoscopes. The overall rate of HCO positivity was 20.4%. Thirty-three resampling procedures were required, and one instrument was temporarily withdrawn from service; no corrective action was necessary in 37 episodes. The most frequently isolated microorganisms were Neisseria subflava (8 isolates) and Micrococcus luteus (7 isolates), followed by coagulase-negative staphylococci (Staphylococcus epidermidis, S. hominis). Pseudomonas aeruginosa, the clinically most relevant pathogen, was isolated in three episodes. Distribution by endoscope type showed higher HCO positivity in duodenoscopes and gastroscopes, whereas colonoscopes (including pediatric models) yielded almost exclusively LCO. Several instruments with≥3 samplings (IDs: 2300238, 2201874, 2200941, 1210792) demonstrated recurrent HCO contamination.
Conclusions
Microbiological monitoring revealed specific criticalities, particularly in duodenoscopes and in instruments exhibiting recurrent HCO positivity. These findings suggest that monitoring should not focus solely on isolated episodes but rather consider the longitudinal performance of each device. Instruments with recurrent HCO contamination may harbor micro-damage or structural defects that facilitate bacterial persistence. In such cases, intensified surveillance, reinforced reprocessing protocols, or instrument replacement may be warranted. Our data confirm the essential role of microbiological monitoring in enhancing the safety of endoscopic procedures and provide insights for future personalized maintenance strategies. Publication History Article published online: 05 June 2026 © 2026. European Society of Gastrointestinal Endoscopy. All rights reserved. Georg Thieme Verlag KG Oswald-Hesse-Straße 50, 70469 Stuttgart, Germany
Zitieren
APA
Dolcini, J., Giacomelli, S., Fiorenza, S., Firmani, G., D'agostino, S., Belvedere, A., Gioia, M. G., Larcinese, L., D'errico, M. M., Barbadoro, P., Montori, M., Bendia, E., & Benedetti, A. (2026). Microbiological Monitoring of Endoscopes: A Single-Center Observational Study. Endoscopy, 58(S 03 S 03), S35. https://doi.org/10.1055/s-0046-1820693
Vancouver
Dolcini J, Giacomelli S, Fiorenza S, Firmani G, D'agostino S, Belvedere A, et al. Microbiological Monitoring of Endoscopes: A Single-Center Observational Study. Endoscopy. 2026;58(S 03 S 03):S35. doi:10.1055/s-0046-1820693
BibTeX
@article{Dolcini2026,
author = {Dolcini, J and Giacomelli, S and Fiorenza, S and Firmani, G and D'agostino, S and Belvedere, A and Gioia, M G and Larcinese, L and D'errico, M M and Barbadoro, P and Montori, M and Bendia, E and Benedetti, A},
title = {Microbiological Monitoring of Endoscopes: A Single-Center Observational Study},
journal = {Endoscopy},
year = {2026},
volume = {58},
number = {S 03 S 03},
pages = {S35},
doi = {10.1055/s-0046-1820693}
}
EUVIMED