Contractility patterns on functional lumen imaging probe (FLIP) prior to peroral endoscopic myotomy predict outcomes in achalasia
Endoscopy · 01.06.2026 · 58 (S 03 S 03) · S12
DOI: 10.1055/s-0046-1820656 ↗Kurz gefasst
Fachartikel in der Zeitschrift Endoscopy, erschienen am 01.06.2026. Verfasst von V Perananthan und 9 weiteren Autor:innen.
Abstract (Originalsprache)
Aims
While functional lumen imaging probe (FLIP) is well-established for identifying esophagogastric junction outflow disorders, its role in patients with confirmed achalasia before per-oral endoscopic myotomy (POEM) are unclear. This study aimed to evaluate whether FLIP provides additional diagnostic value before POEM.
Methods
This retrospective study of prospectively collected data included patients with primary achalasia confirmed on high-resolution esophageal manometry who underwent FLIP at a tertiary center from March 2020 to June 2024. Reduced EGJ opening (REO) on FLIP was defined as a distensibility index 5 cm at 5 minutes and/or retention of a 13-mm barium tablet.
Results
Among 166 patients (44.6% female; median age 60.0, IQR 44.5–73.0), 18 (10.8%) had type I, 123 (74.1%) type II, and 25 (15.1%) had type III achalasia. All patients demonstrated REO on FLIP; 126 (75.9%) had absent CR and 40 (24.1%) had some contractility ([ Table 1 ]). Significant barium retention on TBE was observed in 48 patients (28.9%). Absent CR was associated with increased barium retention (RR 1.3, 95% CI 1.1-1.5, p=0.005) and greater mean column height at 5 minutes (2.0 cm vs. 0.8 cm, p=0.048) post POEM procedure. Table 1 Total cohort No contractility Contractility present p-value N 166 126 (75.9%) 40 (24.1%) Age years (median+IQR) 60.0 (44.5 – 73.0) 61.5 (44.8 – 73.0) 55.0 (44.0 – 73.0) 0.636 Female (n,%) 74 (44.6%) 60 (47.2%) 14 (35.9%) 0.270 BMI (median+IQR) 27.6 (23.6 – 33.0) 26.8 (23.1 – 32.5) 29.9 (26.2 – 34.3) 0.039 * Type 1 (n,%) 18 (10.8%) 14 (11.0%) 4 (10.3%) 0.999 Type 2 (n,%) 123 (74.1%) 98 (77.2%) 25 (64.1%) 0.142 Type 3 (n,%) 25 (15.1%) 15 (11.8%) 10 (25.6%) 0.043 * Pre-POEM Eckardt score (median+IQR) 7.0 (5.0 – 8.0) 7.0 (5.0 – 8.0) 6.0 (4.0 – 9.0) 0.636 Esophageal Myotomy length (cm) 6.0 (6.0 – 8.0) 6.0 (6.0 – 8.0) 8.0 (6.0 – 10.5) 0.114 FLIP Distensibility (mm²/mmHg) at 60mL Pre-POEM (median+IQR) 1.2 (0.8 – 2.3) 1.2 (0.8 – 2.2) 1.2 (0.6 – 2.5) 0.571 FLIP Diameter (mm) at 70mL Pre-POEM (median+IQR) 8.2 (6.2 – 11.5) 8.2 (6.4 – 12.0) 8.7 (5.8 – 10.8) 0.448 Conclusions Absent CR on FLIP prior to POEM predicts significant barium retention on TBE following POEM, supporting an additive role for FLIP to predict outcomes. Publication History Article published online: 05 June 2026 © 2026. European Society of Gastrointestinal Endoscopy. All rights reserved. Georg Thieme Verlag KG Oswald-Hesse-Straße 50, 70469 Stuttgart, Germany
Zitieren
APA
Perananthan, V., Chhoda, A., Velasquez, L. E. O., Song, L. M. W. K., Law, R., Codipilly, D., Snyder, D., Leggett, C., Alexander, J., & Ravi, K. (2026). Contractility patterns on functional lumen imaging probe (FLIP) prior to peroral endoscopic myotomy predict outcomes in achalasia. Endoscopy, 58(S 03 S 03), S12. https://doi.org/10.1055/s-0046-1820656
Vancouver
Perananthan V, Chhoda A, Velasquez LEO, Song LMWK, Law R, Codipilly D, et al. Contractility patterns on functional lumen imaging probe (FLIP) prior to peroral endoscopic myotomy predict outcomes in achalasia. Endoscopy. 2026;58(S 03 S 03):S12. doi:10.1055/s-0046-1820656
BibTeX
@article{Perananthan2026,
author = {Perananthan, V and Chhoda, A and Velasquez, L E Ospina and Song, L M Wong Kee and Law, R and Codipilly, D and Snyder, D and Leggett, C and Alexander, J and Ravi, K},
title = {Contractility patterns on functional lumen imaging probe (FLIP) prior to peroral endoscopic myotomy predict outcomes in achalasia},
journal = {Endoscopy},
year = {2026},
volume = {58},
number = {S 03 S 03},
pages = {S12},
doi = {10.1055/s-0046-1820656}
}
EUVIMED