Evaluation of 3D-Printed Patient-Specific Guides to Facilitate Fluoroscopic-Assisted Kirschner Wire Stabilization of Simulated Femoral Capital Physeal Fractures in 3D-Printed Dog Femur Models
Veterinary and Comparative Orthopaedics and Traumatology · 01.09.2024 · 37 (05) · S111-S142
DOI: 10.1055/s-0044-1791289 ↗Kurz gefasst
Fachartikel in der Zeitschrift Veterinary and Comparative Orthopaedics and Traumatology, erschienen am 01.09.2024. Verfasst von M. Z. Y. Deveci und 6 weiteren Autor:innen.
Abstract (Originalsprache)
Introduction
Three-dimensional-printed (3DP), patient-specific guides have been used to improve accuracy of implant placement in orthopedic procedures. Our objective was to compare speed, accuracy, and ease of Kirschner wire placement for canine capital physeal fracture stabilization between fluoroscopic-assisted freehand placement and the use of 3DP guides in novice and experienced surgeons. Our hypothesis was that 3DP guides would improve Kirschner wire placement by both novice and experienced surgeons. Materials and Methods: Orthopedic surgeons and residents ( n = 4/group) placed three parallel Kirschner wires in five 3DP dog femurs using freehand wire placement and >1 month later using patient-specific 3DP guides, all with fluoroscopic guidance. Post-fixation CT images were compared to preprocedure virtual plans. Outcome measures included procedure time, wire redirections, number of fluoroscopy images, Likert scores (0–10, 10 = very simple), wire insertion and termination locations, and femoral head engagement ratio.
Results
Guided applications reduced the number of fluoroscopy images, wire redirections, and procedure time, while increasing Likert scores ( p < 0.001). Engagement ratios were improved with the guided procedures for the residents ( p < 0.05). Kirschner wires penetrated the articular surface more frequently during freehand applications ( n = 16) than during guide-assisted applications ( n = 4) ( p < 0.01). Discussion/Conclusion: Our study demonstrated that the use of 3DP guides resulted in easier, faster, and more accurate wire placement while reducing fluoroscopy use for both novice and experienced participants during simulated capital physeal fracture stabilization. Reduced procedure time and avoidance of joint penetration with 3DP guide use could improve long-term outcomes. Limitations include simpler guide placement and anatomic reduction with the 3DP bone models. Acknowledgements: We thank those who participated in the study and Dr. Adam Biedrzycki for allowing our use of his 3D printing laboratory. There was no proprietary interest or funding provided for this project. Publication History Article published online: 16 September 2024 Georg Thieme Verlag KG Rüdigerstraße 14, 70469 Stuttgart, Germany
Zitieren
APA
Deveci, M. Z. Y., Worden, N. J., Lewis, D. D., Johnson, M. D., Scheuermann, L. M., Kim, S. E., & Peterson, L. C. (2024). Evaluation of 3D-Printed Patient-Specific Guides to Facilitate Fluoroscopic-Assisted Kirschner Wire Stabilization of Simulated Femoral Capital Physeal Fractures in 3D-Printed Dog Femur Models. Veterinary and Comparative Orthopaedics and Traumatology, 37(05), S111-S142. https://doi.org/10.1055/s-0044-1791289
Vancouver
Deveci MZY, Worden NJ, Lewis DD, Johnson MD, Scheuermann LM, Kim SE, et al. Evaluation of 3D-Printed Patient-Specific Guides to Facilitate Fluoroscopic-Assisted Kirschner Wire Stabilization of Simulated Femoral Capital Physeal Fractures in 3D-Printed Dog Femur Models. Veterinary and Comparative Orthopaedics and Traumatology. 2024;37(05):S111-S142. doi:10.1055/s-0044-1791289
BibTeX
@article{Deveci2024,
author = {Deveci, M. Z. Y. and Worden, N. J. and Lewis, D. D. and Johnson, M. D. and Scheuermann, L. M. and Kim, S. E. and Peterson, L. C.},
title = {Evaluation of 3D-Printed Patient-Specific Guides to Facilitate Fluoroscopic-Assisted Kirschner Wire Stabilization of Simulated Femoral Capital Physeal Fractures in 3D-Printed Dog Femur Models},
journal = {Veterinary and Comparative Orthopaedics and Traumatology},
year = {2024},
volume = {37},
number = {05},
pages = {S111--S142},
doi = {10.1055/s-0044-1791289}
}