Fracture of Aorto-Iliac Stentgraft Delivery System Tip

Charlotte Praca, Arnaud Kerzmann, Evelyne Boesmans, Delphine Szecel, Charlotte Holemans, Jean-Olivier Defraigne

AORTA · 01.10.2023 · 11 (05) · 1-18

DOI: 10.1055/s-0044-1787933 ↗

Kurz gefasst

Fachartikel in der Zeitschrift AORTA, erschienen am 01.10.2023. Verfasst von Charlotte Praca und 5 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Several techniques have been reported to address different endovascular device failures. We report the case of a stentgraft delivery system tip fracture during an endovascular aorto-iliac aneurysm repair. Case report: After deployment of the stentgraft limb, the tip capture mechanism was retracted and the delivery system slowly pulled down. Due to friction on the stentgraft and/or technical defect, the tip of the delivery system broke resulting in a floating foreign body on the wire in the right stentgraft limb. Through the 16 French sheath introduced into the right common femoral artery at the start of the procedure, we dilated the distal part of the right stentgraft limb (figure on the left). By retrieving the balloon, the arterial flow pushed the broken tip into the sheath. The last was removed and the broken tip was pulled out from the sheath (figure on the right) [ Fig. 1 ]. The failure was reported to the company but no clear reason for the tip’s fracture coud be identified.

Conclusion

By dilating the homolateral stentgraft limb, enhanced arterial inflow could push the fractured tip of the stentgraft delivery system into the sheath. Fig. 1 Publication History Article published online: 11 June 2024 © 2024. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution License, permitting unrestricted use, distribution, and reproduction so long as the original work is properly cited. (https://creativecommons.org/licenses/by/4.0/) Thieme Medical Publishers, Inc. 333 Seventh Avenue, 18th Floor, New York, NY 10001, USA

Zitieren

APA

Praca, C., Kerzmann, A., Boesmans, E., Szecel, D., Holemans, C., & Defraigne, J. (2023). Fracture of Aorto-Iliac Stentgraft Delivery System Tip. AORTA, 11(05), 1-18. https://doi.org/10.1055/s-0044-1787933

Vancouver

Praca C, Kerzmann A, Boesmans E, Szecel D, Holemans C, Defraigne J. Fracture of Aorto-Iliac Stentgraft Delivery System Tip. AORTA. 2023;11(05):1-18. doi:10.1055/s-0044-1787933

BibTeX
@article{Praca2023,
  author = {Praca, Charlotte and Kerzmann, Arnaud and Boesmans, Evelyne and Szecel, Delphine and Holemans, Charlotte and Defraigne, Jean-Olivier},
  title = {Fracture of Aorto-Iliac Stentgraft Delivery System Tip},
  journal = {AORTA},
  year = {2023},
  volume = {11},
  number = {05},
  pages = {1--18},
  doi = {10.1055/s-0044-1787933}
}
Kostenlos starten →