EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-SA 4.0

Complication Rate of Percutaneous Balloon- Retention versus Locking-Loop Gastrostomy and Gastrojejunostomy Tube Insertion: A Comparison from a Canadian Tertiary Care Centre

Ian Y. M. Chan, Ibrahim Abdulaziz Alghamdi, Daniel Schep, Sandra Sabongui, Sarah Krause, David Hocking, Daniele Wiseman

The Arab Journal of Interventional Radiology · 01.07.2024 · 08 (02) · 105-111

DOI: 10.1055/s-0044-1786829 ↗

Kurz gefasst

Fachartikel in der Zeitschrift The Arab Journal of Interventional Radiology, erschienen am 01.07.2024. Verfasst von Ian Y. M. Chan und 6 weiteren Autor:innen. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-SA 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

The aim of this study is to compare 30-day complications, procedure-related mortality, and overall mortality rates for de novo enteral feeding tube insertion with fluoroscopy-guided percutaneous balloon-retention versus traditional locking-loop tubes.

Methods

A retrospective analysis was conducted on adult patients who underwent fluoroscopically guided gastrostomy or gastrojejunostomy tube insertions at two tertiary care centers. We categorized complications based on the Society of Interventional Radiology Standards of Practice for Gastrointestinal Access. Factors including the indication for the procedure, the number of gastropexy anchors, and the tube size were analyzed. Statistical analysis was performed using chi-square tests, and the results were compared with patients who underwent locking loop insertions.

Results

A total of 118 patients underwent percutaneous balloon-retention gastrostomy (BRG) or gastrojejunostomy (BRGJ) tube insertions in 2018. These were compared with 559 adult patients who had locking loop insertions at the same institutions from 2011 to 2014. Minor and major complications were higher for the balloon-retention tubes for both BRG (minor: 40.8% vs 4.7%, p

Zitieren

APA

Chan, I. Y. M., Alghamdi, I. A., Schep, D., Sabongui, S., Krause, S., Hocking, D., & Wiseman, D. (2024). Complication Rate of Percutaneous Balloon- Retention versus Locking-Loop Gastrostomy and Gastrojejunostomy Tube Insertion: A Comparison from a Canadian Tertiary Care Centre. The Arab Journal of Interventional Radiology, 08(02), 105-111. https://doi.org/10.1055/s-0044-1786829

Vancouver

Chan IYM, Alghamdi IA, Schep D, Sabongui S, Krause S, Hocking D, et al. Complication Rate of Percutaneous Balloon- Retention versus Locking-Loop Gastrostomy and Gastrojejunostomy Tube Insertion: A Comparison from a Canadian Tertiary Care Centre. The Arab Journal of Interventional Radiology. 2024;08(02):105-111. doi:10.1055/s-0044-1786829

BibTeX
@article{Chan2024,
  author = {Chan, Ian Y. M. and Alghamdi, Ibrahim Abdulaziz and Schep, Daniel and Sabongui, Sandra and Krause, Sarah and Hocking, David and Wiseman, Daniele},
  title = {Complication Rate of Percutaneous Balloon- Retention versus Locking-Loop Gastrostomy and Gastrojejunostomy Tube Insertion: A Comparison from a Canadian Tertiary Care Centre},
  journal = {The Arab Journal of Interventional Radiology},
  year = {2024},
  volume = {08},
  number = {02},
  pages = {105--111},
  doi = {10.1055/s-0044-1786829}
}
Kostenlos starten →