EUVIMEDEUVIMED entdecken
Crossref

Postoperative positional verification of the electrode array using measurements of the local evoked potential in patients with auditory brainstem implants

Lutz Gärtner, Thomas Lenarz, Anke Lesinski-Schiedat, Angelika Illg, Andreas Büchner

Laryngo-Rhino-Otologie · 01.05.2024 · 103 (S 02 S 02) · S254

DOI: 10.1055/s-0044-1784839 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Laryngo-Rhino-Otologie, erschienen am 01.05.2024. Verfasst von Lutz Gärtner und 4 weiteren Autor:innen.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Electrode migration in patients with auditory brainstem implant (ABI) can lead to refusal to use the speech processor, especially in children. Imaging verification of electrode location is costly and also represents radiation exposure. For regular follow-up examinations, measurement of the local evoked potential (LEP) (Gärtner et al. 2021, DOI: 10.1371/journal.pone.0249535) is an alternative. Material/Method Case report: a child fitted with ABI refused to wear the speech processor after osteotomy for relocation of the midface. Measurement of the LEP threshold indicated a slight migration toward the ventricle.

Results

Based on the newly determined LEP thresholds, reprogramming was performed with significantly altered stimulation levels. The program was refined by observing the child's response. Thereafter, the speech processor could be worn again.

Discussion

LEP measurements serve for regular postoperative control of the position of the ABI electrode array and can also be used intraoperatively during ABI surgery for this purpose. The measured LEP thresholds support the adjustment of the speech processor during follow-up. Publication History Article published online: 19 April 2024 © 2024. Thieme. All rights reserved. Georg Thieme Verlag KG Rüdigerstraße 14, 70469 Stuttgart, Germany

Zitieren

APA

Gärtner, L., Lenarz, T., Lesinski-Schiedat, A., Illg, A., & Büchner, A. (2024). Postoperative positional verification of the electrode array using measurements of the local evoked potential in patients with auditory brainstem implants. Laryngo-Rhino-Otologie, 103(S 02 S 02), S254. https://doi.org/10.1055/s-0044-1784839

Vancouver

Gärtner L, Lenarz T, Lesinski-Schiedat A, Illg A, Büchner A. Postoperative positional verification of the electrode array using measurements of the local evoked potential in patients with auditory brainstem implants. Laryngo-Rhino-Otologie. 2024;103(S 02 S 02):S254. doi:10.1055/s-0044-1784839

BibTeX
@article{Grtner2024,
  author = {Gärtner, Lutz and Lenarz, Thomas and Lesinski-Schiedat, Anke and Illg, Angelika and Büchner, Andreas},
  title = {Postoperative positional verification of the electrode array using measurements of the local evoked potential in patients with auditory brainstem implants},
  journal = {Laryngo-Rhino-Otologie},
  year = {2024},
  volume = {103},
  number = {S 02 S 02},
  pages = {S254},
  doi = {10.1055/s-0044-1784839}
}
Kostenlos starten →