A Search Strategy for Detecting Duodenoscope-Associated Infections: A Retrospective Observational Study

K. van der Ploeg, C.H. W. Klaassen, M. C. Vos, J. A. Severin, B.C.G. C. Mason-Slingerland, M. J. Bruno

Endoscopy · 01.04.2024 · 56 (S 02 S 02) · S184-S185

DOI: 10.1055/s-0044-1783104 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Endoscopy, erschienen am 01.04.2024. Verfasst von K. van der Ploeg und 5 weiteren Autor:innen. Die Arbeit nennt 2 Literaturangaben.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Aims

Duodenoscope-associated infections (DAI) are exogenous infections resulting from the use of contaminated duodenoscopes [ 1 ]. While numerous outbreaks of DAI involved multidrug-resistant microorganisms (MDRO), outbreaks involving non-MDRO are also likely to occur [ 2 ]. Detection is hindered because infections are resolved before being cultured or causative strains are not stored for later comparison with strains from contaminated duodenoscopes. This retrospective observational study aims to identify and analyze DAIs that occurred over a seven-year period.

Methods

Duodenoscope cultures positive for gastrointestinal flora between March 2015 and September 2022, were paired with duodenoscope usage data to identify patients exposed to contaminated duodenoscopes. Our analysis encompassed patients treated after a positive duodenoscope culture and those treated within the interval from a negative to a positive culture. Patient identification numbers were cross-referenced with a clinical culture database to identify patients developing infections with matching microorganisms within one year after their procedure. A "pair" was established upon a species-level match between duodenoscope and patient cultures. Pairs were further analyzed via antibiogram comparison, and whole-genome sequencing (WGS) to determine genetic relatedness.

Results

Out of 68 identified pairs, 31 were incomparable due to microorganism unavailability, and 16 exhibited mismatched antibiograms. Twenty-one pairs underwent WGS, revealing three genetically closely related pairs, classified as DAIs. Infection onset occurred up to two months post-procedure. Causative agents included Enterobacter ludwigii , Enterococcus faecium , and Klebsiella pneumoniae , all were non-MDRO.

Conclusions

This study highlights challenges in routine clinical DAI detection. Clinically relevant DAIs with non-MDRO not only exist, but may manifest after a prolonged time following exposure, contributing to the underestimation of DAI risk. Publication History Article published online: 15 April 2024 © 2024. European Society of Gastrointestinal Endoscopy. All rights reserved. Georg Thieme Verlag KG Rüdigerstraße 14, 70469 Stuttgart, Germany

Zitieren

APA

Ploeg, K. V. D., Klaassen, C. W., Vos, M. C., Severin, J. A., Mason-Slingerland, B. C., & Bruno, M. J. (2024). A Search Strategy for Detecting Duodenoscope-Associated Infections: A Retrospective Observational Study. Endoscopy, 56(S 02 S 02), S184-S185. https://doi.org/10.1055/s-0044-1783104

Vancouver

Ploeg KVD, Klaassen CW, Vos MC, Severin JA, Mason-Slingerland BC, Bruno MJ. A Search Strategy for Detecting Duodenoscope-Associated Infections: A Retrospective Observational Study. Endoscopy. 2024;56(S 02 S 02):S184-S185. doi:10.1055/s-0044-1783104

BibTeX
@article{Ploeg2024,
  author = {Ploeg, K. van der and Klaassen, C.H. W. and Vos, M. C. and Severin, J. A. and Mason-Slingerland, B.C.G. C. and Bruno, M. J.},
  title = {A Search Strategy for Detecting Duodenoscope-Associated Infections: A Retrospective Observational Study},
  journal = {Endoscopy},
  year = {2024},
  volume = {56},
  number = {S 02 S 02},
  pages = {S184--S185},
  doi = {10.1055/s-0044-1783104}
}